Background: The purpose was to evaluate the clinical effect of a custom-made Y-shaped fracture fragment reduction device and to assist in posterior unilateral small fenestration of lamina to reduce the fracture fragments.

Methods: In this study, 40 patients were assigned to one of two groups: the traditional reduction device group (TRG) or the Y-shaped reduction device group (YRG). All patients underwent posterior unilateral small fenestration of the lamina and direct decompression through the spinal canal. And the operation time (OT), intraoperative bleeding (IB), preoperative, postoperative, and final follow-up data on the spinal stenosis rate (SSR), Cobb angle, the anterior compression ratio of injured vertebrae (ACRIV), and ASIA neurological function grade were compared between the two groups.

Result: There were no complications, including vascular and nerve injury, serious postoperative infection, internal fixation fracture, or loosening, for any of the patients. And the average follow-up time of the two groups was 14.2 months, the average operation time of the TRG was 236.6 min, and the average intraoperative blood loss was 357.20 ml. Moreover, the average operation time of the YRG was 190.6 min, and the average intraoperative blood loss was 241.5 ml. There were significant differences between the two groups in terms of operation duration and intraoperative blood loss. The YRG's was lower than that of the TRG. Besides, there was no difference in SSR, Cobb angle, ACRIV, or neurological recovery between the two groups before or immediately after the operation or at the last follow-up.

Conclusion: The Y-shaped fracture reduction device can reduce the fracture fragments and the OT and IB stably; it also has satisfactory postoperative curative effects and clinical utility.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10369761PMC
http://dx.doi.org/10.1186/s13018-023-03971-7DOI Listing

Publication Analysis

Top Keywords

reduction device
20
posterior unilateral
12
unilateral small
12
small fenestration
12
fenestration lamina
12
y-shaped fracture
12
operation time
12
intraoperative blood
12
blood loss
12
custom-made y-shaped
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!