For the first time, a paper-based analytical device (PAD) was developed for the assessment of transferrin saturation (TSAT), which is defined as the ratio between iron bound to transferrin (Tf) and the total iron-binding capacity (TIBC) of Tf. Both parameters were simultaneously measured on the same PAD using ferrozine as a chromophore and a smartphone as the color reader. To this end, Tf was first isolated from serum using anti-Tf immunomagnetic beads to ensure that only the Tf-bound iron was measured, improving the selectivity and accuracy of TSAT assessment. To demonstrate the practical utility of the device, it was validated by analyzing a certified reference material, showing excellent accuracy ( < 4%) and good precision (RSD ≤ 6%). Finally, 18 diagnosed serum samples from ischemic stroke patients were analyzed by this approach, and the results were compared with those obtained by urea-PAGE, showing not only an excellent correlation ( = 0.93, < 0.05) but that the PAD approach has become statistically identical to the free-interference urea-PAGE. In comparison with the slow, tedious, and non-miniaturized-PAGE, this PAD approach exhibited attractive characteristics such as low cost, disposability, and connectivity, showing great potential for future testing, especially in developing countries and/or remote areas, where access to medical or clinical facilities is limited.