Aim: To highlight potential benefits of using Reichert's ocular response analyzer (ORA) for intraocular pressure (IOP) measurement following glaucoma filtering surgery (GFS), especially in cases of low IOP.

Background: Goldmann applanation tonometry (GAT) is widely regarded as the gold standard of IOP measurement in clinical practice. The ORA also calculates corneal biomechanics, which is used in the calculation of the cornea-compensated IOP (IOPcc). This useful, previously unknown information can be used to guide management in challenging clinical cases.

Case Description: A 78-year-old lady underwent right trabeculectomy with mitomycin C. During the first 9 months postoperatively, IOP's when measured with GAT, were found to be low (ranging between 2 mm Hg and 5 mm Hg). The patient displayed no clinical features of hypotony and visual acuity remained stable throughout. IOP was also measured with Reichert's ORA, which gives readings for corneal hysteresis (CH), corneal resistance factor and corneal corrected IOP. IOPcc measurements obtained with ORA were consistently higher, ranging from 6.7 mm Hg to 9.3 mm Hg and were more in keeping with the clinical features of the case.

Conclusion: This case highlights the possibility that GAT can underestimate the true IOP in eyes with low IOP following GFS, as it does not account for corneal biomechanical properties.

Clinical Significance: Ocular response analyzer (ORA)-measured IOPcc may be a useful adjunct in reassuring surgeons to manage postoperative numerical hypotony conservatively in the absence of any clinical signs of hypotony.

How To Cite This Article: Dawson S, Mohite A. Use of Ocular Response Analyzer Output Data in the Management of Low Intraocular Pressure after Glaucoma Filtration Surgery. J Curr Glaucoma Pract 2023;17(2):104-105.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10357027PMC
http://dx.doi.org/10.5005/jp-journals-10078-1402DOI Listing

Publication Analysis

Top Keywords

ocular response
16
response analyzer
16
intraocular pressure
12
analyzer output
8
output data
8
data management
8
management low
8
low intraocular
8
pressure glaucoma
8
glaucoma filtration
8

Similar Publications

Pyrimidinergic P2Y1-Like Nucleotide Receptors Are Functional in Rat Conjunctival Goblet Cells.

Invest Ophthalmol Vis Sci

January 2025

Schepens Eye Research Institute, Massachusetts Eye and Ear Infirmary, Department of Ophthalmology, Harvard Medical School, Boston, Massachusetts, United States.

Purpose: To investigate the presence of uridine-5'-triphosphate (UTP)-activated P2Y1-like nucleotide receptors (P2Y2R, P2Y4R, and P2Y6R) in conjunctival goblet cells (CGCs) and determine if they increase intracellular Ca2+ concentration ([Ca2+]i) and induce mucin secretion.

Methods: Adult, male rat conjunctiva was used for culture of CGCs. To investigate the expression of P2YRs, mRNA was extracted from CGCs and used for reverse transcription PCR (RT-PCR) with commercially obtained primers specific to P2Y2R, P2Y4R, and P2Y6R.

View Article and Find Full Text PDF

Introduction: Dry eye disease (DED) can impact the accuracy of biometry measurements prior to cataract surgery (CS), influence visual performance post-CS, and can be exacerbated by CS. We performed a survey to evaluate the DED practice of clinicians directly caring for CS patients.

Design: Prospective face-to-face survey.

View Article and Find Full Text PDF

Purpose: Inconsistency exists among reported studies on the composition of human ocular surface microbiome (OSM). The roles of OSM in ocular diseases remain uncertain. In this study, we aimed to determine the composition of OSM and to evaluate its potential roles and functions from multiple cohorts.

View Article and Find Full Text PDF

External environments (e.g., pollutants, irritants, ultraviolet radiation, etc) probably activate oxidative stress on the ocular surface, further leading to inflammatory responses and cellular apoptosis.

View Article and Find Full Text PDF

Hyperosmotic stress-induced NLRP3 inflammasome activation via the mechanosensitive PIEZO1 channel in dry eye corneal epithelium.

Ocul Surf

January 2025

Eye hospital and School of Ophthalmology and Optometry, Wenzhou Medical University, Zhejiang, 325000, China; State Key Laboratory of Ophthalmology,Optometry and Visual Science, Wenzhou Medical University,Zhejiang, 325000,China. Electronic address:

Unlabelled: The activation of the NLRP3 inflammasome by hyperosmotic stress is a critical pathophysiological response in dry eye disease (DED), driving the chronic cycle of inflammation on the ocular surface. The specific mechanism underlying hyperosmotic mechanical stimulation activates the NLRP3 inflammasome remains unclear. This study provides evidence that PIEZO1, a mechanosensitive ion channel, functions as the primary receptor for corneal epithelial cells in sensing mechanical stimulation induced by tear hyperosmolarity.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!