Objectives: The present in vitro study aimed to evaluate the accuracy of three-dimensional (3D) printed indirect bonding trays consisting of hard or soft resin materials produced using computer-aided design and manufacturing (CAD/CAM).
Methods: Forty-eight dental casts were 3D printed. Four groups based on frontal crowding were defined and divided into hard- and soft-resin groups. After virtual bracket positioning on the digital models, the transfer trays were 3D printed. To evaluate the accuracy of the procedure, measurements were performed using a digital overlay of the virtual (target) bracket position and a post-bonding scan. The horizontal, transverse, and vertical deviations and angular discrepancies were analyzed. The loss rate was evaluated descriptively as a percentage.
Results: A total of 553 brackets were bonded using 24 soft and 24 resilient indirect bonding trays. The mean deviations were of 0.05 mm (transversal), 0.05 mm (horizontal), 0.09 mm (vertical), 0.13° (angulation) in the resilient resin group and of 0.01 mm (transversal), 0.08 mm (horizontal), 0.08 mm (vertical), 0.37° (angular) in the soft resin group. The loss rate was 6.9% and 0.7% in the hard and soft resin groups, respectively. Angular deviations were significantly higher in the soft resin group (P = 0.009), whereas the loss rate was considerably higher in the hard resin group (P < 0.001).
Significance: The findings indicate that indirect bonding using CAD/CAM is an accurate procedure in the laboratory setting. Soft resins are considered favorable for loss rate and useability.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.dental.2023.07.003 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!