Background: Preprocedural computed tomography (CT) workup with assessment of virtual transcatheter heart valve-to-coronary ostia (VTC) distance and transcatheter heart valve-to-sinus (VTS) distances is recommended to assess the risk of coronary obstruction following valve-in-valve (ViV) transcatheter aortic valve replacement (TAVR).

Objectives: The authors sought to investigate the agreement of predicted VTC and VTS distances and observed post-TAVR anatomy on CT and their relationship with transcatheter heart valve (THV) expansion and deployment conditions.

Methods: Fifty-one patients who underwent a balloon-expandable ViV procedure were included in this study. The expansion of the THV stent frame was evaluated at 4 levels: THV inflow, surgical heart valve (SHV) sewing ring, SHV outflow, and THV outflow. Assessment of the VTC/VTS distances was performed on the pre-TAVR CT, and THV-to-coronary ostia and THV-to-sinus distances were assessed on the post-TAVR CT.

Results: Following the ViV procedure, the THV stent frame flared toward the outflow but was generally underexpanded at all levels, particularly at the SHV sewing ring level. Postdilatation impacted the extent of THV expansion, resulting in greater expansion than nominal balloon filling at all 4 THV levels (P < 0.001). Observed THV-to-coronary ostia distances were systematically larger than predicted by the VTC distance (mean difference 1.25 ±1.28 mm) in patients with nominal balloon filling but systematically smaller in case of postdilatation (mean difference -0.45 ± 0.52 mm). A similar relationship was observed between VTS and THV-to-sinus distance measurements.

Conclusions: With nominal balloon filling, VTC and VTS distances underestimate postprocedural distances due to THV frame underexpansion. However, postdilatation may lead to distances smaller than predicted due to THV overexpansion at the outflow level.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcin.2023.05.038DOI Listing

Publication Analysis

Top Keywords

transcatheter heart
12
vts distances
12
nominal balloon
12
balloon filling
12
thv
9
computed tomography
8
vtc distance
8
distances
8
predicted vtc
8
vtc vts
8

Similar Publications

Background: Despite advancements in valve implantation devices, vascular access complications (VAC) remain a significant cause of morbidity and mortality for those undergoing transcatheter aortic valve replacement (TAVR). We describe pre-operative imaging analysis of the aortoiliac and femoral arterial beds using the TransAtlantic intersociety consensus (TASC) score, ilio-femoral tortuosity, and procedural characteristics to identify anatomic risk factors predictive of VAC in TAVR.

Methods: Consecutive patients undergoing TAVR from 2012 to 2022 at a single North Dakota hospital were retrospectively reviewed.

View Article and Find Full Text PDF

Background: Transcatheter aortic valve replacement (TAVR) is a well-established treatment option for patients with severe aortic valve stenosis; however, clinical valve thrombosis is a major challenge.

Case Summary: A 92-year-old woman underwent TAVR for severe aortic stenosis. One month later, the patient developed acute heart failure.

View Article and Find Full Text PDF

Background: Right ventricular (RV) function assessment by echocardiography can be challenging due to its complex morphology. Also, increasing use of sedation rather than general anesthesia for transfemoral approach transcatheter aortic valve replacement (TAVR) reduces the need for intraoperative transesophageal echocardiography (TEE). Recent clinical studies have demonstrated the importance of 3-dimensional (3D) echocardiography and a longitudinal strain for RV function assessment.

View Article and Find Full Text PDF

Acute Decompensated Valvular Disease in the Intensive Care Unit.

JACC Adv

December 2024

TIMI Study Group, Cardiovascular Division, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA.

Acute decompensated valvular disease encompasses a group of complex and challenging conditions, which are often the primary reason for admission to the cardiac intensive care unit and can also complicate the management of other primary cardiac disorders. Critically ill patients with valvular disease also present unique diagnostic and management challenges. Historically, medical and percutaneous interventional therapies have been limited and surgery was the only definitive treatment; however, surgical risk can at times be prohibitive.

View Article and Find Full Text PDF

Transcatheter caval valve implantation (CAVI) is an emerging treatment option for inoperable patients with severe tricuspid regurgitation (TR). We present a case of a 76-year-old lady with severe TR and recurrent right heart failure who improved after CAVI. This is the first such case in a rare environment.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!