Background: Highly localized impedance (LI) measurements during atrial fibrillation (AF) ablation have the potential to act as a reliable predictor of the durability of the lesions created.

Objective: We aimed to collect data on the procedural parameters affecting LI-guided ablation in a large multicenter registry.

Methods: A total of 212 consecutive patients enrolled in the CHARISMA registry and undergoing their first pulmonary vein (PV) isolation for paroxysmal and persistent AF were included.

Results: In all, 13,891 radiofrequency (RF) applications of ≥3 s duration were assessed. The first-pass PV isolation rate was 93.3%. A total of 80 PV gaps were detected. At successful ablation spots, baseline LI and absolute LI drop were larger than at PV gap spots (161.4 ± 19 Ω vs. 153.0 ± 13 Ω,  < 0.0001 for baseline LI; 22.1 ± 9 Ω vs. 14.4 ± 5 Ω,  < 0.0001 for LI drop). On the basis of Receiver operating characteristic curve analysis, the ideal LI drop, which predicted successful ablation, was >21 Ω at anterior sites and >18 Ω at posterior sites. There was a non-linear association between the magnitude of LI drop and contact-force (CF) ( = 0.14, 95% CI: 0.13-0.16,  < 0.0001) whereas both CF and LI drop were inversely related with delivery time (DT) (-0.22, -0.23 to -0.20,  < 0.0001 for CF; -0.27, -0.29 to -0.26,  < 0.0001 for LI drop).

Conclusion: An LI drop >21 Ω at anterior sites and >18 Ω at posterior sites predicts successful ablation. A higher CF was associated with an increased likelihood of ideal LI drop. The combination of good CF and adequate LI drop allows a significant reduction in RF DT.

Clinical Trial Registration: http://clinicaltrials.gov/, identifier: NCT03793998.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10354239PMC
http://dx.doi.org/10.3389/fcvm.2023.1169037DOI Listing

Publication Analysis

Top Keywords

pulmonary vein
8
vein isolation
8
insight contact
4
contact force
4
force local
4
local impedance
4
impedance technology
4
technology predicting
4
predicting effective
4
effective pulmonary
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!