Background: Airway fistula is a rare but threatening complication associated with high rates of morbidity and mortality. We report the experience of Amplatzer device application in airway fistulae that failed to be cured with a covered self-expandable metallic stent (SEMS).
Materials And Methods: Patients who failed occlusion with a covered self-expandable metallic stent and received Amplatzer device placement from Jan 2015 to Jan 2020 were retrospectively enrolled. A total of 14 patients aged 42 to 66 years (55.14 ± 7.87) were enrolled in this study. The primary diseases, types of fistula, types of stents, duration, size of fistula, and follow-up were recorded.
Results: All 14 patients with airway fistula failed to be occluded with a covered metallic stent and received Amplatzer device placement. Among the 14 patients, 6 had BPF, 3 had TEF and 5 had GBF. The average stent time was 141.93 ± 65.83 days. The sizes of the fistulae ranged from 3 to 6 mm. After Amplatzer device placement, the KPS score improved from 62.14 ± 4.26 to 75.71 ± 5.13 (P < 0.05). No procedure-related complications occurred. During the 1-month, 3-month and 6-month follow-ups, all the Amplatzer devices were partially surrounded with granulation. Only 1 patient with BPF failed with Amplatzer device occlusion due to the recurrence of lung cancer.
Conclusion: In conclusion, the application of the Amplatzer device is a safe and effective option in the treatment of airway fistula that failed to be occluded with SEMSs.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10357874 | PMC |
http://dx.doi.org/10.1186/s12890-023-02548-8 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!