Purpose: To evaluate computationally the optical performance of AcrySof IQ Vivity extended depth-of-focus (EDOF) intraocular lenses (IOLs) in post-laser in situ keratomileusis (LASIK) eyes.
Setting: Visual Optics and Biophotonics Laboratory, Madrid, Spain.
Design: Experimental study.
Methods: Computer pseudophakic eye models were implemented using reported post-LASIK corneal aberrations (refractive corrections from -7.5 to +4.5 diopters [D]) and virtually implanted with monofocal (AcrySof IQ) or EDOF (AcrySof IQ Vivity) IOLs. Retinal image quality was quantified through visual Strehl (VS). The depth of focus (DOF) was calculated from the through-focus VS curves. Halos were estimated from the light spread in the image of a pinhole. Those quantitative parameters were obtained for 5.0 and 3.0 mm pupil diameters.
Results: Simulated virgin eyes showed VS of 0.89/0.99 with monofocal IOLs and 0.74/0.52 with EDOF IOLs for 5.0/3.0 mm pupils at best focus. VS decreased with induced spherical aberration (SA) by 25% and with induced SA + coma by 61% on average (3.0 mm pupils). The DOF was 2.50 D in virgin eyes with EDOF IOLs, 1.66 ± 0.30 and 2.54 ± 0.31 D ( P < .05) on average in post-LASIK eyes for 3.0 mm pupils, monofocal and EDOF IOLs, respectively. Halos were more sensitive to SA induction for 5.0 mm pupils, and induction of positive SA (myopic LASIK) resulted in reduced halos with the EDOF when compared with the monofocal IOLs, by 1.62 (SA) and 1.86 arc min (SA + coma), on average.
Conclusions: Computer post-LASIK pseudophakic eye models showed that the DOF was less dependent on the presence of SA and coma with EDOF IOLs and that halos were reduced with EDOF IOLs compared with the monofocal IOL for a range of SA.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/j.jcrs.0000000000001260 | DOI Listing |
Vestn Oftalmol
December 2024
Krasnov Research Institute of Eye Diseases, Moscow, Russia.
Purpose: This study compares the functional outcomes of correction using two different types of aberrational extended depth-of-focus (EDOF) intraocular lenses (IOLs).
Material And Methods: The study was conducted in two standardized groups (20 patients, 20 eyes in each group). Patients in group 1 were implanted EDOF IOL I, in group 2 - EDOF IOL II.
Biomed Opt Express
December 2024
Laboratorio de Óptica, Universidad de Murcia, Campus de Espinardo, 30100 Murcia, Spain.
Cataract surgery involves the implantation of an intraocular lens (IOL) to replace the opacified crystalline lens. Monofocal IOLs, the most common type, are intended to have the eye in focus at a given distance, usually at infinity. Simultaneous vision IOLs (SVIOLs) and extended depth of focus (EDOF) aim to minimize postoperative dependence on spectacles by providing either multiple foci or an extended depth of focus.
View Article and Find Full Text PDFBr J Ophthalmol
December 2024
Department of Ophthalmology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea
Biomed Opt Express
November 2024
Institute of Optics, Spanish National Research Council, IO-CSIC, Madrid, Spain.
This study presents a systematic method to simulate various intraocular lenses (IOLs) available in the market. Five IOLs (two trifocals, one bifocal, one enhanced monofocal, and one extended depth of focus (EDOF)) were evaluated in terms of through focus visual Strehl (TFVS) utilizing the OptiSpheric IOL PRO2 device (Trioptics GmbH). Then, the estimated TFVS (ETFVS) and the temporal coefficients necessary for temporal multiplexing were computed, and through an iterative process, the SimVis TFVS was obtained.
View Article and Find Full Text PDFInt Ophthalmol
November 2024
Department of Ophthalmology, Graduate School of Medicine and Pharmaceutical Sciences, University of Toyama, 2630 Sugitani, Toyama City, 930-0194, Japan.
Purpose: We investigated the effect of enhanced monofocal intraocular lenses (IOL) (TECNIS Eyhance DIB00V) on retinal sensitivity in glaucoma patients after cataract surgery by comparing results with of such patients implanted with non-enhanced monofocal intraocular lenses (TECNIS ZCB00V).
Patients And Methods: This was a multi-facility retrospective study of glaucoma patients who had undergone cataract surgery. We measured the value of the foveal threshold with a Humphrey visual field analyzer before and after cataract surgery and considered this value to represent the sensitivity of the retina.