Purpose: To investigate the clinical validity of using wavefront measurements obtained with a recently available pyramidal aberrometer to assess the optical quality of eyes implanted with diffractive intraocular lenses (IOLs).
Methods: Individual biometric data were used to create models of pseudophakic eyes implanted with two diffractive IOLs. Their synthetic wavefronts were calculated by ray-tracing with near infrared wavelength (0.85 μm). Comparisons of the through-focus visual acuity of 12 pseudophakic eyes were obtained with three different methods: clinical defocus curves; simulated defocus curves calculated from ray-tracing in the customized model eyes; and through-focus simulated defocus curves calculated from the wavefront data measured with a pyramidal aberrometer.
Results: Image quality calculated from wavefront data obtained by ray-tracing with 0.85 μm wavelength, without scaling the phase to 0.55 μm, resulted in a significantly different through-focus curve compared to the reference values. Even so, after scaling of the wavefront data to 0.55 μm, the defocus curves calculated from the wavefronts measured with the pyramidal aberrometer did not match the shape and the depth of field of the clinical defocus curves or the theoretical expected values.
Conclusions: Correcting for the longitudinal chromatic aberration of the eye when measuring the wavefront of eyes implanted with diffractive IOLs under near infrared light only accounts for the best focus shift due to the longitudinal chromatic aberration, but not for the wavelength dependence of the diffractive element. The pyramidal sensor does not seem to properly sample the slopes of a wavefront measured from a pseudophakic eye implanted with a presbyopia-correcting diffractive IOL to a clinically acceptable level. .
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.3928/1081597X-20230523-01 | DOI Listing |
BMC Ophthalmol
January 2025
Department of Ophthalmology, Eye & ENT Hospital, Fudan University, 83 Fenyang Road, Xuhui District, Shanghai, 200031, China.
Background: To investigate the impact of central corneal astigmatism on postoperative visual outcomes in patients with trifocal intraocular lens (IOL) implantation.
Methods: This retrospective study included 278 eyes of 278 patients who underwent uneventful cataract surgery with implantation of the trifocal IOL (AT LISA tri 839MP). Patients were divided into two groups according to the total corneal refractive power (TCRP) in 3 mm zone centered on the corneal apex: low astigmatism group, TCRP ≤ 0.
Klin Monbl Augenheilkd
January 2025
Department of Ophthalmology, Pallas Kliniken, Olten/Bern/Zürich/Dübendorf, Switzerland.
Background: Extended monovision is a novel mix-and-match approach that has been recently introduced. It involves implanting an aspherical monofocal intraocular lens (IOL) for distance vision in the dominant eye, and a bifocal extended depth-of-focus (EDOF) IOL in the nondominant eye. The target refraction for the nondominant eye is - 1.
View Article and Find Full Text PDFPurpose: To observe and explore the correlation between visual outcomes and intraocular lens (IOL) stability after tri-focal IOL implantation in eyes with high myopia.
Methods: Patients with highly myopic cataract (axial length > 26 mm) were enrolled in this prospective study. Thirty-one eyes (31 patients) received implantation of a trifocal IOL (AcrySof IQ PanOptix TFNT00).
Diagnostics (Basel)
December 2024
Department of Ophthalmology, Asan Medical Center, University of Ulsan College of Medicine, Seoul 05505, Republic of Korea.
Objectives: This study compared the visual outcomes and optical quality of two monofocal, aspheric intraocular lenses (IOLs; CT LUCIA 621P, Carl Zeiss Meditec; Eyhance ICB00, Johnson & Johnson Vision) by evaluating visual acuity, contrast sensitivity, and higher-order aberrations 1 month post-cataract surgery.
Methods: In this retrospective, comparative study, 120 eyes (72 patients) that underwent cataract surgery with either CT LUCIA 621P (Lucia group) or Eyhance ICB00 (Eyhance group) implantation (60 eyes/group) were retrospectively investigated. Visual acuity at various distances and defocus curves were measured 1 month postoperatively.
Clin Ophthalmol
December 2024
Département de Recherche Clinique, Institut Ophtalmologique de l'Ouest (IOO) Jules Verne, Nantes, France.
Purpose: To evaluate clinical outcomes obtained after cataract surgery involving the implantation of a trifocal hydrophobic intraocular lens (IOL) and to determine if pupil size and the corneal aberrometric profile correlate to visual acuity at different distances.
Methods: 49 patients (98 eyes) underwent bilateral cataract surgery with the placement of FineVision HP IOLs for presbyopia and were assessed at 1- and 3- to 6-months post-surgery. Postoperatively, refraction, monocular and binocular uncorrected and corrected distance visual acuity (UDVA and CDVA), uncorrected intermediate visual acuity (UIVA), uncorrected near visual acuity (UNVA) and the binocular defocus curve were measured.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!