Ahmed to Baerveldt glaucoma drainage device exchange in pediatric patients.

BMC Ophthalmol

Department of Ophthalmology, Northwestern University Feinberg School of Medicine, 645 N. Michigan Ave, Chicago, IL, 60611, USA.

Published: July 2023

Background: There is no consensus and few reports as to the surgical management of encapsulated Ahmed glaucoma drainage devices (GDD) which no longer control intraocular pressure (IOP), especially within the pediatric population. The purpose of this study was to report outcomes of exchanging the Ahmed GDD for a Baerveldt GDD in children with refractory glaucoma.

Methods: Retrospective review of children (< 18yrs) who underwent removal of Ahmed FP7 and placement of Baerveldt 350 (2016-2021) with ≥ 3-month follow-up. Surgical success was defined as IOP 5-20 mmHg without additional IOP-lowering surgeries or visually devastating complications. Outcomes included change in best-corrected visual acuity (BCVA), intraocular pressure (IOP), and number of glaucoma medications.

Results: Twelve eyes of 10 patients underwent superotemporal Ahmed FP7 to Baerveldt 350 GDD exchange at 8.8 ± 3.6 years. Time to Ahmed failure was 2.7 ± 1.9 years with 1-, 3-, and 5-year survival rates of 83% with a 95% CI[48,95], 33% with a 95% CI[10, 59], and 8% with a 95% CI[0, 30]. At final follow-up (2.5 ± 1.8 years), success rate for Baerveldt 350 GDDs was 75% (9 of 12 eyes) with 1 and 3-yr survival rates of 100% and 71% with 95% CI[25,92], respectively. IOP (24.1 ± 2.9 vs. 14.9 ± 3.1 mmHg) and number of glaucoma medications (3.7 ± 0.7 vs. 2.7 ± 1.1) were significantly decreased (p < 0.004). BCVA remained stable. Two eyes required cycloablation and 1 eye developed a retinal detachment.

Conclusions: Ahmed removal with Baerveldt placement can improve IOP control with fewer medications in cases of refractory pediatric glaucoma. However, more eyes with greater follow-up are required to determine long-term outcomes.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10334620PMC
http://dx.doi.org/10.1186/s12886-023-03074-1DOI Listing

Publication Analysis

Top Keywords

glaucoma drainage
8
ahmed baerveldt
4
baerveldt glaucoma
4
drainage device
4
device exchange
4
exchange pediatric
4
pediatric patients
4
patients background
4
background consensus
4
consensus reports
4

Similar Publications

Purpose: Pressure resistance characteristics of the Paul glaucoma implant (PGI) and Ahmed ClearPath 250 (ACP), with and without the insertion of polypropylene thread in their tubes, were evaluated.

Methods: The in vitro flow pressure was evaluated at varying flow rates, both with and without threads (6-0 for PGI and 4-0 or 3-0 for ACP). Cross-sectional areas of the tube lumen and thread were measured to calculate pressure resistance using the Hagen-Poiseuille equation.

View Article and Find Full Text PDF

Is minimally invasive glaucoma surgery killing trabeculectomies?

Curr Opin Ophthalmol

November 2024

Centre for Vision Research, Westmead Institute for Medical Research, University of Sydney.

Purpose Of Review: The introduction of minimally invasive glaucoma surgery (MIGS) has changed the surgical landscape of glaucoma surgery. For more than four decades, the most commonly performed glaucoma surgery around the world has been the trabeculectomy. This article examines whether trabeculectomy is still the predominant procedure in glaucoma today, in the context of rapidly emerging MIGS devices and procedures.

View Article and Find Full Text PDF

Background: Addressing oculoplastic conditions in the preoperative period ensures both the safety and functional success of any ophthalmic procedure. Some oculoplastic conditions, like nasolacrimal duct obstruction, have been extensively studied, whereas others, like eyelid malposition and thyroid eye disease, have received minimal or no research.

Aim: To investigate the current practice patterns among ophthalmologists while treating concomitant oculoplastic conditions before any subspecialty ophthalmic intervention.

View Article and Find Full Text PDF

Objectives: Femtosecond laser image guided high precision trabeculotomy (FLigHT) is a novel open-angle glaucoma treatment. The procedure non-invasively creates aqueous humor (AH) drainage channel from the anterior chamber (AC) to Schlemm's canal (SC) through the trabecular meshwork (TM) to decrease intraocular pressure (IOP). The purpose of this study was to develop a 3D finite element model (FEM) of the FLigHT procedure and to simulate clinical results for different drainage channel cross-sectional areas.

View Article and Find Full Text PDF

Schlemm's canal endothelial cells (SECs) serve as the final barrier to aqueous humor (AQH) drainage from the eye. SECs adjust permeability to AQH outflow to modulate intraocular pressure (IOP). The broad identification of IOP-related genes implicates SECs in glaucoma.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!