Background: The impact of filtering on bipolar electrograms (EGMs) has not been systematically examined. We tried to clarify the optimal filter configuration for ventricular tachycardia (VT) ablation.

Methods: Fifteen patients with VT were included. Eight different filter configurations were prospectively created for the distal bipoles of the ablation catheter: 1.0-250, 10-250, 100-250, 30-50, 30-100, 30-250, 30-500, and 30-1000 Hz. Pre-ablation stable EGMs with good contact (contact force > 10 g) were analyzed. Baseline fluctuation, baseline noise, bipolar peak-to-peak voltage, and presence of local abnormal ventricular activity (LAVA) were compared between different filter configurations.

Results: In total, 2276 EGMs with multiple bipolar configurations in 246 sites in scar and border areas were analyzed. Baseline fluctuation was only observed in the high-pass filter of (HPF) ≤ 10 Hz (p < .001). Noise level was lowest at 30-50 Hz (0.018 [0.012-0.029] mV), increased as the low-pass filter (LPF) extended, and was highest at 30-1000 Hz (0.047 [0.041-0.061] mV) (p < .001). Conversely, the HPF did not affect the noise level at ≤30 Hz. As the HPF extended to 100 Hz, bipolar voltages significantly decreased (p < .001), but were not affected when the LPF was extended to ≥100 Hz. LAVAs were most frequently detected at 30-250 Hz (207/246; 84.2%) and 30-500 Hz (208/246; 84.6%), followed by 30-1000 Hz (205/246; 83.3%), but frequently missed at LPF ≤ 100 Hz or HPF ≤ 10 Hz (p < .001). A 50-Hz notch-filter reduced the bipolar voltage by 43.9% and LAVA-detection by 34.5% (p < .0001).

Conclusion: Bipolar EGMs are strongly affected by filter settings in scar/border areas. In all, 30-250 or 30-500 Hz may be the best configuration, minimizing the baseline fluctuation, baseline noise, and detecting LAVAs. Not applying the 50-Hz notch filter may be beneficial to avoid missing VT substrate.

Download full-text PDF

Source
http://dx.doi.org/10.1111/jce.15997DOI Listing

Publication Analysis

Top Keywords

filter configurations
8
optimal filter
8
analyzed baseline
8
baseline fluctuation
8
filter
5
impact filter
4
bipolar
4
configurations bipolar
4
egms
4
bipolar egms
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!