First results from the ACURATE Prime XL human feasibility study.

Cardiovasc Revasc Med

Boston Scientific Corporation, Marlborough, MA, USA. Electronic address:

Published: December 2023

Background/purpose: This prospective, open-label, single-arm study evaluated transcatheter aortic valve replacement (TAVR) in patients with severe aortic stenosis with ACURATE Prime XL, an iteration of the ACURATE neo2 device designed with improved radial force and adaptations for compatibility with a larger annulus diameter (≥ 26.5 mm and ≤ 29 mm based on pre-procedure diagnostic imaging).

Methods: The composite primary device success endpoint was based on Valve Academic Research Consortium (VARC)-2 criteria. The primary safety endpoint was a composite of all-cause mortality and all stroke at 30 days. Aortic valve (AV) performance, including mean AV gradient, AV area, and grade of paravalvular leak (PVL), was assessed by an independent core laboratory.

Results: 13 male patients were enrolled at 3 Australian centers (mean age: 83.1 years; 10 of 13 were considered high/extreme operative risk). 61.5 % of patients met the primary device success endpoint. At 30 days, no patients experienced death or stroke; one patient received a permanent pacemaker. Mean AV gradient improved from baseline (42.7 ± 11.0 mmHg) to discharge (7.7 ± 2.5 mmHg) and through 30 days (7.2 ± 2.3 mmHg). Mean AV area was 0.8 ± 0.1 cm at baseline, 1.9 ± 0.3 cm at discharge, and 1.7 ± 0.3 cm at 30 days. Per core-laboratory adjudication, no patient had moderate or severe PVL at 30 days; 91.7 % had no/trace PVL and 8.3 % had mild PVL.

Conclusions And Relevance: In this first-in-human feasibility study of the ACURATE Prime XL valve, there were no safety concerns, and no deaths or strokes within 30 days. Valve hemodynamics were favorable, and no patient had >mild PVL.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.carrev.2023.06.028DOI Listing

Publication Analysis

Top Keywords

acurate prime
12
feasibility study
8
aortic valve
8
primary device
8
device success
8
success endpoint
8
30 days
6
valve
5
acurate
4
prime human
4

Similar Publications

Background: There are limited data regarding treatment for failed balloon-expandable transcatheter heart valves (THVs) in redo-transcatheter aortic valve implantation (TAVI).

Aims: We aimed to assess THV performance, neoskirt height and expansion when performing redo-TAVI with the ACURATE platform inside a SAPIEN 3 (S3) compared to redo-TAVI with an S3 in an S3.

Methods: Redo-TAVI was performed on the bench using each available size of the S3, the ACURATE neo2 (ACn2) and the next-generation ACURATE Prime XL (AC XL) implanted at 2 different depths within 20 mm/23 mm/26 mm/29 mm S3s serving as the "failed" index THV.

View Article and Find Full Text PDF

First results from the ACURATE Prime XL human feasibility study.

Cardiovasc Revasc Med

December 2023

Boston Scientific Corporation, Marlborough, MA, USA. Electronic address:

Background/purpose: This prospective, open-label, single-arm study evaluated transcatheter aortic valve replacement (TAVR) in patients with severe aortic stenosis with ACURATE Prime XL, an iteration of the ACURATE neo2 device designed with improved radial force and adaptations for compatibility with a larger annulus diameter (≥ 26.5 mm and ≤ 29 mm based on pre-procedure diagnostic imaging).

Methods: The composite primary device success endpoint was based on Valve Academic Research Consortium (VARC)-2 criteria.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!