Background: Most of the scientific societies recommend assessing the accuracy of electronic devices for blood pressure (BP) measurements using established validation protocol.

Objective: To determine the accuracy of the BP measurements using the Withings BPM Core device in the general population according to the "Universal Standard (ISO 81060-2:2018/AMD 1:2020)".

Methods: The Withings BPM Core is an oscillometric device measuring BP at the brachial level. The study was performed according to the "Universal Standard (ISO 81060-2:2018/AMD 1:2020) protocol" using the same-arm sequential BP measurement method. Subjects (n ≥ 85) fulfilling the age, gender, BP, and cuff distribution criteria of the protocol were included. Analysis was performed as required by the Universal protocol using Criterion 1 - differences between observers' mercury sphygmomanometer reference measurements and test device BP values (test versus reference) and their standard deviation (SD); and Criterion 2 - The SD of the mean BP differences between the test device and reference BP per subject.

Results: Eighty-six subjects were selected, 85 of whom were included. The mean BP differences between the simultaneous two observers' measurements were -0.2 ± 2.1 mmHg for systolic BP (SBP) and 0.3 ± 2.1 mmHg for diastolic BP (DBP). For validation criterion 1, the mean difference ± SD between the reference and device BP values was -0.6 ± 4.8 mmHg for SBP and 0.1 ± 3.7 mmHg for DBP (≤5 ± 8 mmHg for both SBP and DBP). For criterion 2, the SD of the mean BP differences between the test device and reference BP per subject was 3.2/2.6 mmHg for SBP and DBP (≤6.91/6.95 mmHg).

Conclusion: The results of this study showed that the Withings BPM Core oscillometric device for home BP measurement fulfilled the accuracy requirements of the (ISO 81060-2:2018/AMD 1:2020) Universal protocol in the general population.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10328101PMC
http://dx.doi.org/10.2147/VHRM.S413195DOI Listing

Publication Analysis

Top Keywords

withings bpm
16
bpm core
16
general population
12
iso 81060-22018/amd
12
criterion differences
12
test device
12
mmhg sbp
12
device
8
core device
8
pressure measurements
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!