AI Article Synopsis

  • The SCOPE I trial compared the safety and efficacy of two transcatheter aortic valve devices: the self-expanding ACURATE Neo (NEO) and the balloon-expandable SAPIEN 3 (S3), finding that NEO had higher rates of prosthetic valve regurgitation and acute kidney injury at 30 days.
  • A follow-up analysis involving 739 patients found that after 3 years, clinical outcomes such as mortality, stroke, and hospitalizations were similar between groups, suggesting early performance differences did not impact long-term results.
  • Overall, the study concludes that both valves have comparable outcomes over three years, despite initial concerns with the NEO device.

Article Abstract

Background: In the SCOPE I trial (Safety and Efficacy of the Symetis ACURATE Neo/TF Compared to the Edwards SAPIEN 3 Bioprosthesis), transcatheter aortic valve implantation with the self-expanding ACURATE neo (NEO) did not meet noninferiority compared with the balloon-expandable SAPIEN 3 (S3) device regarding a composite end point at 30 days due to higher rates of prosthetic valve regurgitation and acute kidney injury. Data on long-term durability of NEO are scarce. Here, we report whether early differences between NEO and S3 translate into differences in clinical outcomes or bioprosthetic valve failure 3 years after transcatheter aortic valve implantation.

Methods: Patients with severe aortic stenosis were randomized to transfemoral transcatheter aortic valve implantation with NEO or S3 at 20 European centers. Clinical outcomes at 3 years are compared using Cox proportional or Fine-Gray subdistribution hazard models by intention-to-treat. Bioprosthetic valve failure is reported for the valve-implant cohort.

Results: Among 739 patients, 84 of 372 patients (24.3%) had died in the NEO and 85 of 367 (25%) in the S3 group at 3 years. Comparing NEO with S3, the 3-year rates of all-cause death (hazard ratio, 0.98 [95% CI, 0.73-1.33]), stroke (subhazard ratio, 1.04 [95% CI, 0.56-1.92]), and hospitalization for congestive heart failure (subhazard ratio, 0.74 [95% CI, 0.51-1.07]) were similar between the groups. Aortic valve reinterventions were required in 4 NEO and 3 S3 patients (subhazard ratio, 1.32 [95% CI, 0.30-5.85]). New York Heart Association functional class ≤II was observed in 84% (NEO) and 85% (S3), respectively. Mean gradients remained lower after NEO at 3 years (8 versus 12 mm Hg; <0.001).

Conclusions: Early differences between NEO and S3 did not translate into significant differences in clinical outcomes or bioprosthetic valve failure throughout 3 years.

Registration: URL: https://clinicaltrials.gov, Unique identifier: NCT03011346.

Download full-text PDF

Source
http://dx.doi.org/10.1161/CIRCINTERVENTIONS.123.012873DOI Listing

Publication Analysis

Top Keywords

aortic valve
20
transcatheter aortic
16
subhazard ratio
12
neo
10
valve
8
valve implantation
8
clinical outcomes
8
bioprosthetic valve
8
valve failure
8
aortic
6

Similar Publications

Osteogenic differentiation is crucial in normal bone formation and pathological calcification, such as calcific aortic valve disease (CAVD). Understanding the proteomic and transcriptomic landscapes underlying this differentiation can unveil potential therapeutic targets for CAVD. In this study, we employed RNA sequencing transcriptomics and proteomics on a timsTOF Pro platform to explore the multiomics profiles of valve interstitial cells (VICs) and osteoblasts during osteogenic differentiation.

View Article and Find Full Text PDF

Transcatheter versus Surgical Aortic Valve Replacement in Patients Younger than 65 Years in the US.

J Thorac Cardiovasc Surg

January 2025

Department of Cardiac Surgery, Smidt Heart Institute, Cedars-Sinai Medical Center, Los Angeles, California, USA. Electronic address:

Objective: To characterize trends and outcomes of aortic valve replacement in patients <65 with aortic stenosis between 2013 and 2021.

Methods: This retrospective analysis included 9,557 patients who underwent biological aortic valve replacement in California, New York, and New Jersey from 2013 through 2021. Patients were stratified by approach: transcatheter aortic valve replacement (TAVR) versus surgical aortic valve replacement (SAVR).

View Article and Find Full Text PDF

Aortic valve leaflet assessment to inform novel bioinspired materials: Understanding the impact of collagen fibres on the tissue's mechanical behaviour.

J Mech Behav Biomed Mater

December 2024

Trinity Centre for Biomedical Engineering, Trinity Biomedical Sciences Institute, Trinity College Dublin, 2, Dublin, Ireland; Discipline of Mechanical, Manufacturing, and Biomedical Engineering, School of Engineering, Trinity College Dublin, 2, Dublin, Ireland; Advanced Materials and Bioengineering Research Centre (AMBER), Trinity College Dublin, Ireland. Electronic address:

Aortic stenosis is a prevalent disease that is treated with either mechanical or bioprosthetic valve replacement devices. However, these implants can experience problems with either functionality in the case of mechanical valves or long-term durability in the case of bioprosthetic valves. To enhance next generation prosthetic valves, such as biomimetic polymeric valves, an improved understanding of the native aortic valve leaflet structure and mechanical response is required to provide much needed benchmarks for future device development.

View Article and Find Full Text PDF
Article Synopsis
  • The study focuses on the outcomes of aortic valve replacement (AVR) in young Moroccan patients with severe aortic regurgitation and impaired left ventricle function.
  • A total of 42 out of 110 patients with significant aortic regurgitation and an ejection fraction of ≤35% underwent AVR, showing a low immediate postoperative mortality rate of 4.7%.
  • Long-term follow-up indicated a high survival rate of 86.8% after 15 years, along with significant clinical improvements and increased left ventricle ejection fraction post-surgery.
View Article and Find Full Text PDF