Introduction: The aims of lens removal surgery are to re-establish or preserve both a clear visual axis and emmetropic vision. Trans-scleral intraocular lens (IOL) fixation has been described in cases where lens capsule instability precludes the insertion of a prosthetic intraocular lens into the lens capsule. Previous techniques have necessitated enlargement of the corneal incision to accommodate either a rigid polymethylmethacrylate IOL or an acrylic foldable IOL inserted using forceps. This paper reports the modification of an endocapsular IOL to be used as an injectable suture-fixated IOL introduced through a 2.8 mm corneal incision.

Materials And Methods: All cases underwent lens extraction by phacoemulsification followed by removal of the unstable lens capsule. A PFI X4 IOL (Medicontur) was modified to create four open-loop haptics. The IOL was injected into the anterior chamber, each haptic was captured in a loop of suture introduced ab externo, and the lens was sutured with four-point fixation.

Results: The results from 20 eyes in 17 dogs are reported. Over an average follow-up time of 14.5 months, vision was retained in 16/20 eyes. Vision was lost in four eyes due to corneal ulceration and ocular hypertension (1/20), retinal detachment (2/20), and Progressive Retinal Atrophy (1/20).

Conclusions: The modified PFI X4 proved suitable for injection and scleral fixation through a 2.8 mm corneal incision, with a success rate comparable to previously published techniques.

Download full-text PDF

Source
http://dx.doi.org/10.1111/vop.13129DOI Listing

Publication Analysis

Top Keywords

intraocular lens
12
lens capsule
12
lens
9
scleral fixation
8
corneal incision
8
28 mm corneal
8
iol
7
fixation novel
4
novel modified
4
modified injected
4

Similar Publications

Background: Large language models have shown remarkable efficacy in various medical research and clinical applications. However, their skills in medical image recognition and subsequent report generation or question answering (QA) remain limited.

Objective: We aim to finetune a multimodal, transformer-based model for generating medical reports from slit lamp images and develop a QA system using Llama2.

View Article and Find Full Text PDF

Purpose Of The Review: In recent years, Berger's space (BS), the potential space between the posterior lens capsule and the anterior hyaloid membrane, has received little attention from the ophthalmic clinical community. This is primarily due to the limited documentation, with only a few isolated case reports detailing foreign bodies in this area.

Recent Findings: Recent advances in medical imaging technology have enabled the visualization of the BS under various circumstances.

View Article and Find Full Text PDF

Purpose: To evaluate long-term anatomical and functional outcomes in eyes undergoing scleral fixation of aniridic black diaphragm intraocular lens (BDIOL).

Methods: Total 38 eyes that underwent scleral fixation of BDIOL for secondary aniridia following ocular trauma with aphakia with absent posterior capsule or primary congenital aniridia with subluxated lens were retrospectively analyzed. Final visual acuity, early and late complications were examined.

View Article and Find Full Text PDF

Photic phenomena are more pronounced in presbyopia-corrected than in monofocal intraocular lens (IOL), causing dissatisfaction after cataract surgery. Photic Phenomena Test (PPT) quantifies photic phenomena in eyes with two types of presbyopia-corrected IOL. We examined the relationship between preoperative eye shape and pupil diameter.

View Article and Find Full Text PDF

A lensless compact arrangement based on digital in-line holography under Gabor's regime is proposed as a novel contactless method to assess the profile of multifocal intraocular lenses (MIOLs) which are conformed by several diffractive rings. Diffractive MIOLs are a widely adopted ophthalmologic option for the correction of presbyopia in patients undergoing cataract surgery. The MIOL optical design might introduce non-negligible optical performance differences between lenses as well as the introduction of undesirable photic phenomena (such as halos and glare) affecting the vision of users.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!