Purpose: To investigate the accuracy of posterior chamber phakic intraocular lens (PIOL) vault and size prediction models based on sulcus to sulcus (STS) optimized artificial intelligence and big data analysis technology.
Design: Big data and artificial intelligence prediction model.
Methods: We included 5873 eyes with posterior chamber PIOL implantation, and the postoperative vault was measured using an anterior segment analyzer (Pentacam AXL) 1 month postoperatively. A random forest regression model and classification model were used to predict the postoperative vault and PIOL size. The postoperative vault and PIOL size were set as output features; other vault-related eye parameters were set as input features. The influence of white to white (WTW), horizontal sulcus to sulcus (STS), and vertical STS on predicting postoperative vault and PIOL size was analyzed and compared.
Results: The mean preoperative WTW diameter was 11.64 ± 0.37 mm, the mean horizontal STS diameter was 11.85 ± 0.47 mm, and the mean vertical STS diameter was 12.39 ± 0.52 mm. In the regression model for numerical prediction of the vault, the combination of WTW, horizontal STS, and vertical STS was the most optimal for vault prediction (R = 0.3091, root mean square error [RMSE] = 0.1705); solely relying on WTW was the least optimal (R = 0.2849, RMSE = 0.1735). Among the models for classification prediction of the vault, the combination of WTW, horizontal STS, and vertical STS was the most accurate (accuracy, 0.6302; mean area under the curve, 0.8008; and mean precision recall rate, 0.6940). Moreover, the combination of WTW, horizontal STS, and vertical STS exhibited the highest accuracy for classification prediction of PIOL size (accuracy, 0.8170; mean area under the curve, 0.9540; and mean precision recall rate, 0.8864). Whether in the regression prediction models of vault values or in the classification prediction models of vault and PIOL size, the accuracy of STS optimized model was significantly improved compared with the traditional WTW model (P < .001).
Conclusion: Artificial intelligence combined with STS optimization contributes to the accuracy of PIOL size and vault prediction models. The random forest machine-learning model optimized by STS is superior to the traditional WTW model.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ajo.2023.06.024 | DOI Listing |
Front Med (Lausanne)
August 2024
Shenzhen Eye Institute, Shenzhen Eye Hospital, Jinan University, Shenzhen, China.
BMC Ophthalmol
July 2024
Department of Ophthalmology, Hangil Eye Hospital, 35 Bupyeong-Daero, Bupyeong-Gu, Incheon, 21388, South Korea.
Purpose: Measurement of corneal endothelial cells is critical for postoperative evaluation of phakic intraocular lens (pIOL) surgery. However, inter-instrument differences in corneal endothelial cell density (ECD) after pIOL implantation have not yet been reported. This study aimed to compare automated corneal endothelial cell analysis between CellChek-20 (Konan Medical, Hyogo, Japan) and EM-4000 (Tomey, Nagoya, Japan) in healthy and postoperative eyes with pIOL.
View Article and Find Full Text PDFFront Med (Lausanne)
April 2024
Department of Ophthalmology, The Third People's Hospital of Dalian, Dalian, China.
Introduction: Myopia is causing a major public health concern, with its prevalence increasing globally. This study aimed to discuss posterior chamber phakic intraocular lens (pIOL) research publication trends and hotspots over the past 20 years.
Methods: Bibliometric analysis was performed using the Web Science Core Collection to investigate posterior-chamber pIOL research publication trends.
Clin Ophthalmol
April 2024
Eye Institute and Department of Ophthalmology, Eye & ENT Hospital, Fudan University, Shanghai, People's Republic of China.
Zhonghua Yan Ke Za Zhi
December 2023
Department of Ophthalmology, Peking University Third Hospital, Beijing Key Laboratory of Restoration of Damaged Ocular Nerve, Beijing 100191, China.
Posterior chamber phakic intraocular lens (pIOL) implantation has been widely adopted for the correction of refractive errors. Among pIOLs, the Implantable Collamer Lens is the most common choice. The selection of the appropriate pIOL size and achieving the desired postoperative vault to minimize complications has consistently been a focal point in academic research.