Inter-assay variation between different immunoassays and different mass spectrometry methods hampers the biochemical confirmation of male hypogonadism. Furthermore, some laboratories utilis eassay manufacturer reference ranges that do not necessarily mirror assay performance characteristics, with the lower limit of normality ranging from 4.9 nmol/L to 11 nmol/L. The quality of the normative data underlying commercial immunoassay reference ranges is uncertain. A working group reviewed published evidence and agreed upon standardised reporting guidance to augment total testosterone reports. Evidence-based guidance on appropriate blood sampling, clinical action limits, and other major factors likely to affect the interpretation of results are provided. This article aims to improve the quality of the interpretation of testosterone results by non-specialist clinicians. It also discusses approaches for assay harmonisation which have been successful in some but not all healthcare systems.