Background: Ventricular pacing can cause myocardial dysfunction, but how lead anchoring to the myocardium affects function has not been studied.
Objective: The purpose of this study was to evaluate patterns of regional and global ventricular function in patients with a ventricular lead using cine cardiac computed tomography (CCT) and histology.
Methods: This was a single-center retrospective study with 2 groups of patients with a ventricular lead: (1) those who underwent cine CCT from September 2020 to June 2021 and (2) those whose cardiac specimen was analyzed histologically. Regional wall motion abnormalities on CCT were assessed in relation to lead characteristics.
Results: For the CCT group, 122 ventricular lead insertion sites were analyzed in 43 patients (47% female; median age 19 years; range 3-57 years). Regional wall motion abnormalities were present at 51 of 122 lead insertion sites (42%) in 23 of 43 patients (53%). The prevalence of a lead insertion-associated regional wall motion abnormality was higher with active pacing (55% vs 18%; P < .001). Patients with lead insertion-associated regional wall motion abnormalities had a lower systemic ventricular ejection fraction (median 38% vs 53%; P < .001) than did those without regional wall motion abnormalities. For the histology group, 3 patients with 10 epicardial lead insertion sites were studied. Myocardial compression, fibrosis, and calcifications were commonly present directly under active leads.
Conclusion: Lead insertion site-associated regional wall motion abnormalities are common and associated with systemic ventricular dysfunction. Histopathological alterations including myocardial compression, fibrosis, and calcifications beneath active leads may explain this finding.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.hrthm.2023.06.018 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!