Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Spot-check hemoglobin co-oximetry analyzers measure hemoglobin transcutaneously and offer the benefit of a hemoglobin measurement without phlebotomy. The objective of this study was to determine the validity of non-invasive spot-check hemoglobin co-oximetry testing for the detection of postpartum anemia (hemoglobin < 10 g/dL).
Methods: Five hundred eighty-four women aged 18 and over were recruited on postpartum day one following a singleton delivery. Two non-invasive spot-check hemoglobin co-oximetry monitors, Masimo Pronto Pulse CO-Oximeter (Pronto) and Masimo Rad-67 Pulse CO-Oximeter (Rad-67), were evaluated and compared to the postpartum phlebotomy hemoglobin value.
Results: Of 584 participants, 31% (181) had postpartum anemia by phlebotomy hemoglobin measurement. Bland-Altman plots determined a bias of + 2.4 (± 1.2) g/dL with the Pronto and + 2.2 (± 1.1) g/dL with the Rad-67. Low sensitivity was observed: 15% for the Pronto and 16% for the Rad-67. Adjusting for the fixed bias, the Pronto demonstrated a sensitivity of 68% and specificity of 84%, while the Rad-67 demonstrated a sensitivity of 78% and specificity of 88%.
Conclusion: A consistent overestimation of hemoglobin by the non-invasive spot-check hemoglobin co-oximetry monitors compared to phlebotomy hemoglobin result was observed. Even after adjusting for the fixed bias, the sensitivity for detecting postpartum anemia was low. Detection of postpartum anemia should not be based on these devices alone.