Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 143
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 143
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 209
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3098
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 574
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 488
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Severity: Warning
Message: Attempt to read property "Count" on bool
Filename: helpers/my_audit_helper.php
Line Number: 3100
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3100
Function: _error_handler
File: /var/www/html/application/controllers/Detail.php
Line: 574
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 488
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objective: The Woven EndoBridge (WEB) was developed to treat wide-neck bifurcation intracranial aneurysms. Occasionally, persistent contrast-filling has been observed in the WEB after treatment. The purpose of our study was to investigate its incidence, predictive factors, and clinical impact.
Methods: All patients treated with the WEB between January 2021 and September 2021 at our institute were reviewed. Age, gender, antiplatelet therapy, and angioarchitecture were compared between the persistent-filling group and the no-filling group at the three-month follow-up angiography.
Results: We included 20 patients with 20 unruptured aneurysms. Ten of the 20 intracranial aneurysms (50 %) showed contrast-filling in the WEB after three months. Two of the 10 had contrast not only inside, but around the device. Statistically significant differences were observed between the persistent-filling group and the no-filling group in neck size (median: 4.5 mm vs. 3.8 mm), deviation of the aneurysm axis from the inlet flow line where the orifice of the bifurcated arteries overlaps (mean: 15.1° vs. 33.0°), and postoperative dual antiplatelet therapy (DAPT) for at least a month (90 % vs. 20 %). One case had additional coil embolization six months after the WEB implantation. Including this case, one year after the treatment or the re-treatment, the filling in the device had disappeared in nine of 10 cases. No bleeding was observed during the mean follow-up period of 24 months.
Conclusion: Persistent contrast-filling was associated with postoperative DAPT for at least a month, a wide neck, and less deviation of the aneurysm axis from the inlet flow line. If the contrast-filling is only within the WEB and not between the aneurysmal wall and the WEB, we are not concerned. To further assess the clinical impact of this phenomenon, long-term follow-up will be needed.