Introduction: The efficacy and safety of leadless cardiac pacemakers (LPMs) as an alternative to conventional transvenous cardiac pacing have been largely reported. The first generation of the Micra transcatheter pacing system (VR; Medtronic) was able to provide single-chamber VVI(R) pacing mode only, with a potential risk of pacemaker syndrome in sinus rhythm patients. A second-generation system (AV) now provides atrioventricular synchrony through atrial mechanical (Am) sensing capability (VDD mode).
Objective: We sought to compare VR and AV systems in sinus rhythm patients with chronic ventricular pacing (Vp) for complete atrioventricular block.
Methods: All consecutive patients implanted with an LPM in our department for complete atrioventricular block were retrospectively screened. Patients with atrial fibrillation, sinus dysfunction, or Vp burden <20% at 1 month postimplantation were excluded. Patients were systematically followed with a visit at 1 month, and then at least once a year.
Results: A total of 93 patients-45 VR (2015-2020) and 48 AV (2020-2021)-were included. VR and AV patients had similar baseline characteristics, except for VR patients being older (80 ± 8 vs. 77 ± 9 years, p = 0.049). The mean Vp burden was 77% in the VR and 82% in the AV group (p = 0.38). In AV patients, the median AV synchronous beats rate was 78%, with 65% having a >66% rate. An E/A ratio <1.2 as measured on echocardiography was the only independent predictor of accurate atrial mechanical tracking (p = 0.01). One-year survival rate was similar in both groups. Five patients in the VR and 0 in the AV group eventually developed pacemaker syndrome within 1 year post-implantation (p = 0.02).
Conclusion: In sinus rhythm patients with chronic Vp for complete atrioventricular block implanted with an LPM, the atrial mechanical sensing algorithm allowed significant atrioventricular synchrony in most patients and was associated with no occurrence of-otherwise rare-pacemaker syndrome.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/jce.15981 | DOI Listing |
Sci Rep
December 2024
Division of Cardiology, Department of Internal Medicine, The Jikei University School of Medicine, 3-25-8 Nishi-shimbashi, Minato-ku, Tokyo, 105-8461, Japan.
B-type natriuretic peptide (BNP) levels accurately reflect the degree of cardiac overload in heart failure. Considering cardiac morphology and intracardiac pressure, including the left ventricular end-systolic volume index (LVESVI) and left ventricular end-diastolic volume index (LVEDVI), is essential for cardiac overload assessment. These indexes influence plasma BNP levels, and high heart rate is likely associated with cardiac morphology.
View Article and Find Full Text PDFFront Cardiovasc Med
December 2024
Arrhythmology, Pacing and Electrophysiology Unit, Cardiology Service, Santa Marta Hospital, Lisbon, Portugal.
Introduction: Low atrial voltage and slow conduction velocity (CV) have been associated with atrial fibrillation (AF); however, their interaction and relative importance as early disease markers remain incompletely understood. We aimed to elucidate the relationship between atrial voltage and CV using high-density electroanatomic (HDE) maps of patients with AF.
Methods: HDE maps obtained during sinus rhythm in 52 patients with AF and five healthy controls were analysed.
BMC Med Educ
December 2024
Department of Pediatrics and Child Health, School of Medicine, College of Health Sciences, Addis Ababa University, Addis Ababa, Ethiopia.
Background: Electrocardiography (ECG) interpretation competency is vital to ensure the timely initiation of life-saving treatment for emergent ECG conditions. This competency has not been well-studied among pediatric and child health residents. Hence, the study aimed to determine the competency in ECG interpretation and its predictors among residents at the National University of Ethiopia.
View Article and Find Full Text PDFFront Pharmacol
December 2024
Department of Pharmacy, Zhengzhou No. 7 People's Hospital, Zhengzhou, Henan, China.
Background: Studies have shown that DNA methylation of the CACNA1C gene is involved in the pathogenesis of various diseases and the mechanism of drug action. However, its relationship with atrial fibrillation (AF) remains largely unexplored.
Objective: To investigate the association between DNA methylation of the CACNA1C gene and AF by combining decitabine (5-Aza-2'-deoxycytidine, AZA) treatment with multi-omics analysis.
BMC Cardiovasc Disord
December 2024
Department of Electrocardiology, The Third Affiliated Hospital of Wenzhou Medical University, No.108 WansongRoad, Wenzhou, 325200, People's Republic of China.
Background: Ventricular separation is a multipart, extensive disease of the heart that hinders the electrical conduction of the cardiac system ventricular muscle, causing a bidirectional conduction block. The occurrence of ventricular separation suggests that the myocardium is in a state of severe ischemia, and the prognosis is generally poor. Herein, we present arescue case in which the extremely rare phenomenon of ventricular separation developed and was documented in realtime.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!