An preliminary investigation into the impact of parameters on tissue welding strength in small intestine mucosa-mucosa end-to-end anastomosis.

Front Bioeng Biotechnol

Department of Light Sources and Illuminating Engineering, School of Information Science and Technology, Fudan University, Shanghai, China.

Published: June 2023

Tissue welding is an electrosurgical technique that can fuse tissue for small intestine anastomosis. However, limited knowledge exists on its application in mucosa-mucosa end-to-end anastomosis. This study investigates the effects of initial compression pressure, out-put power, and duration time on anastomosis strength in mucosa-mucosa end-to-end anastomosis. porcine bowel segments were used to create 140 mucosa-mucosa end-to-end fusions. Different experimental parameters were employed for fusion, including initial com-pression pressure (50kPa-400 kPa), output power (90W, 110W, and 140W), and fusion time (5, 10, 15, 20 s). The fusion quality was measured by burst pressure and optical microscopes. The best fusion quality was achieved with an initial compressive pressure between 200 and 250 kPa, an output power of 140W, and a fusion time of 15 s. However, an increase in output power and duration time resulted in a wider range of thermal damage. There was no significant difference between the burst pressure at 15 and 20 s ( > 0.05). However, a substantial increase in thermal damage was observed with longer fusion times of 15 and 20 s ( < 0.05). The best fusion quality for mucosa-mucosa end-to-end anastomosis is achieved when the initial compressive pressure is between 200 and 250 kPa, the output power is approximately 140W, and the fusion time is approximately 15 s. These findings can serve as a valuable theoretical foundation and technical guidance for conducting animal experiments and subsequent tissue regeneration.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10277648PMC
http://dx.doi.org/10.3389/fbioe.2023.1200239DOI Listing

Publication Analysis

Top Keywords

mucosa-mucosa end-to-end
20
end-to-end anastomosis
16
output power
16
140w fusion
12
fusion time
12
fusion quality
12
tissue welding
8
small intestine
8
power duration
8
duration time
8

Similar Publications

An preliminary investigation into the impact of parameters on tissue welding strength in small intestine mucosa-mucosa end-to-end anastomosis.

Front Bioeng Biotechnol

June 2023

Department of Light Sources and Illuminating Engineering, School of Information Science and Technology, Fudan University, Shanghai, China.

Tissue welding is an electrosurgical technique that can fuse tissue for small intestine anastomosis. However, limited knowledge exists on its application in mucosa-mucosa end-to-end anastomosis. This study investigates the effects of initial compression pressure, out-put power, and duration time on anastomosis strength in mucosa-mucosa end-to-end anastomosis.

View Article and Find Full Text PDF