Objective: Drug-resistant epilepsy (DRE) can have devastating consequences for patients and families. Vagal nerve stimulation (VNS) is used as a surgical adjunct for treating DRE not amenable to surgical resection. Although VNS is generally safe, it has its inherent complications. With the increasing number of implantations, adequate patient education with discussion of possible complications forms a critical aspect of informed consent and patient counseling. There is a lack of large-scale reviews of device malfunction, patient complaints, and surgically related complications available to date.

Materials And Methods: Complications associated with VNS implants performed between 2011 and 2021 were identified through a search of the United States Food and Drug Administration Manufacturer And User Facility Device Experience (MAUDE) data base. We found three models on the data base, CYBERONICS, INC pulse gen Demipulse 103, AspireSR 106, and SenTiva 1000. The reports were classified into three main groups, "Device malfunction," "Patient complaints," and "Surgically managed complications."

Results: A total of 5888 complications were reported over the ten-year period, of which 501 reports were inconclusive, 610 were unrelated, and 449 were deaths. In summary, there were 2272 reports for VNS 103, 1526 reports for VNS 106, and 530 reports for VNS 1000. Within VNS 103, 33% of reports were related to device malfunction, 33% to patient complaints, and 34% to surgically managed complications. For VNS 106, 35% were related to device malfunction, 24% to patient complaints, and 41% to surgically managed complications. Lastly, for VNS 1000, 8% were device malfunction, 45% patient complaints, and 47% surgically managed complications.

Conclusion: We present an analysis of the MAUDE data base for adverse events and complications related to VNS. It is hoped that this description of complications and literature review will help promote further improvement in its safety profile, patient education, and management of both patient and clinician expectations.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.neurom.2023.04.474DOI Listing

Publication Analysis

Top Keywords

device malfunction
16
patient complaints
16
data base
12
reports vns
12
surgically managed
12
complications
10
vns
10
adverse events
8
events complications
8
complications associated
8

Similar Publications

Reel's syndrome (RS) is an unusual cause of pacemaker lead dislodgement. We present the case of a 59-year-old female patient with Down syndrome (DS) implanted with a dual-chamber endovascular pacemaker due to symptomatic sinus node disfunction, reporting several syncopal episodes in last days and showing abnormal electrical parameters at the 2-months follow-up due to RS. The malfunctioning device was removed and an endocardial leadless pacing system was implanted.

View Article and Find Full Text PDF

Rare case of cochlear implant damaged by nearby use of monopolar electrosurgery.

Am J Otolaryngol

December 2024

Department of Otolaryngology-Head & Neck Surgery, University of California San Francisco, San Francisco, CA, United States of America. Electronic address:

Objectives: To present a rare case of a cochlear implant (CI) damaged by nearby use of monopolar electrosurgery.

Patient: A 38-year-old man with a right-sided CI reported that his implant had stopped producing sound immediately after his meningioma resection.

Interventions: Right pterional craniotomy with use of monopolar electrosurgery.

View Article and Find Full Text PDF

The growing prevalence of coronary artery diseases in the US corresponds to the increasing use of minimally invasive techniques that require coronary stents. Although extensive research is available on the perioperative outcomes of the 3 stent options - bare-metal stents (BMS), drug-eluting stents (DES), and bioresorbable drug-eluting stents (BVS), a knowledge gap exists in the longitudinal monitoring of patient outcomes due to device-related causes. Therefore, our study examines the device-related patient outcome and the relative performance for BMS, DES, and BVS.

View Article and Find Full Text PDF

Background: Endoscopic bariatric interventions are viable alternatives to traditional weight loss surgeries for patients with a body mass index between 30 and 40 kg/m. While studies have explored the impact of intragastric balloons on obesity and weight reduction, comprehensive data on real-world device-related complications and failures are limited. This study aimed to investigate mechanical failures leading to patient adverse events reported to the US Food and Drug Administration.

View Article and Find Full Text PDF