AI Article Synopsis

  • The study aimed to assess the effectiveness of a quality assurance (QA) program based on the American College of Radiology's (ACR) CT quality control manual to evaluate a clinical photon-counting detector (PCD) CT system.
  • A daily QA program was implemented to test CT number accuracy and artifacts, examining both standard and ultra-high-resolution scan modes, while several tools and methods, like the modulation transfer function (MTF), were used for comprehensive evaluations.
  • The findings revealed that while CT number accuracy varied with image types, proper protocols and the use of the 70 keV virtual monoenergetic image (VMI) allowed compliance with ACR standards, highlighting the need for

Article Abstract

Purpose: To determine the suitability of a quality assurance (QA) program based on the American College of Radiology's (ACR) CT quality control (QC) manual to fully evaluate the unique capabilities of a clinical photon-counting-detector (PCD) CT system.

Methods: A daily QA program was established to evaluate CT number accuracy and artifacts for both standard and ultra-high-resolution (UHR) scan modes. A complete system performance evaluation was conducted in accordance with the ACR CT QC manual by scanning the CT Accreditation Phantom with routine clinical protocols and reconstructing low-energy-threshold (T3D) and virtual monoenergetic images (VMIs) between 40 and 120 keV. Spatial resolution was evaluated by computing the modulation transfer function (MTF) for the UHR mode, and multi-energy performance was evaluated by scanning a body phantom containing four iodine inserts with concentrations between 2 and 15 mg I/cc.

Results: The daily QA program identified instances when the detector needed recalibration or replacement. CT number accuracy was impacted by image type: CT numbers at 70 keV VMI were within the acceptable range (defined for 120 kV). Other keV VMIs and the T3D reconstruction had at least one insert with CT number outside the acceptable range. The limiting resolution was nearly 40 lp/cm based on MTF measurements, which far exceeds the 12 lp/cm maximum capability of the ACR phantom. The CT numbers in the iodine inserts were accurate on all VMIs (3.8% average percentage error), while the iodine concentrations had an average root mean squared error of 0.3 mg I/cc.

Conclusion: Protocols and parameters must be properly selected on PCD-CT to meet current accreditation requirements with the ACR CT phantom. Use of the 70 keV VMI allowed passing all tests prescribed in the ACR CT manual. Additional evaluations such an MTF measurement and multi-energy phantom scans are also recommended to comprehensively evaluate PCD-CT scanner performance.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10338779PMC
http://dx.doi.org/10.1002/acm2.14074DOI Listing

Publication Analysis

Top Keywords

quality assurance
8
assurance program
8
daily program
8
number accuracy
8
acr manual
8
iodine inserts
8
70 kev vmi
8
acceptable range
8
acr phantom
8
acr
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!