Irritant contact dermatitis (ICD) occurs frequently with the use of diabetes devices, but no guidelines for treatment exist. Since subsequent devices need intact skin for intended use, quick healing is crucial. Normal wound healing is expected to be 7-10 days. This was a single-center cross-over study that investigated the effectiveness of an occlusive hydrocolloid-based patch versus nonocclusive treatment of ICD. Participants were aged 6-20 years with active ICD caused by using diabetes device. First study period was patch treatment for 3 days. A control arm was initiated if a new ICD occurred within 30 days. ICD healed completely in 21% of the patch group but none in the controls. Itching was reported as an adverse event (AE) in both arms, but only one additional AE was noted in the patch arm: an infection at a different site from investigated. The hydrocolloid-based patch showed signs of faster healing of ICD with no additional AEs, but larger studies are needed.

Download full-text PDF

Source
http://dx.doi.org/10.1089/dia.2023.0224DOI Listing

Publication Analysis

Top Keywords

hydrocolloid-based patch
12
occlusive hydrocolloid-based
8
irritant contact
8
contact dermatitis
8
diabetes devices
8
patch
6
icd
6
patch effective
4
effective feasible
4
feasible safe
4

Similar Publications