Rapid diagnostic testing for syphilis in Arctic communities (the STAR study): a multisite prospective field diagnostic accuracy study in an intended-use setting.

Clin Microbiol Infect

Infectious Diseases and Immunity in Global Health, Research Institute of the McGill University Health Centre, Montreal, Quebec, Canada; McGill Interdisciplinary Initiative in Infection and Immunity, Montreal, Quebec, Canada; J.D. MacLean Centre for Tropical Diseases, McGill University, Montreal, Quebec, Canada; Divisions of Infectious Diseases and Medical Microbiology, McGill University Health Centre, Montreal, Quebec, Canada. Electronic address:

Published: October 2023

Objectives: We evaluated the field diagnostic accuracy of a syphilis rapid test (RDT), using serum and whole blood by non-laboratorians in two Canadian Arctic communities.

Methods: We implemented a multisite prospective field evaluation wherein patients were screened by an RDT containing treponemal and non-treponemal components (Chembio DPP® Syphilis Screen & Confirm) between January 2020 and December 2021. Venous whole blood and serum were collected for rapid testing and compared with laboratory-based serology reference testing using a reverse sequence algorithm of treponemal and rapid plasma reagin (RPR) testing.

Results: Overall, 135 whole blood and 139 serum specimens were collected from 161 participants during clinical encounters. Treponemal-RDT sensitivity against a treponemal-reference standard (38/161 confirmed cases) was similar for serum (78% [95% CI: 61-90%]) and whole blood (81% [95% CI: 63-93%]). In those with RPR titres ≥1:8 (i.e. suggestive of recent/active infection), sensitivity increased to 93% (95% CI: 77-99%) for serum and 92% (95% CI: 73-99%) for whole blood. Treponemal-RDT specificity was excellent (99% [95% CI: 95-100%]) for both specimen types. Non-treponemal-RDT sensitivity against RPR was 94% (95% CI: 80-99%) for serum and 79% (95% CI: 60-92%) for whole blood. Sensitivity increased to 100% (95% CI: 88-100%) for serum and 92% (95% CI: 73-99%) for whole blood when RPR titres ≥1:8. RDT performance with whole blood was similar to that with serum.

Discussion: Non-laboratorians using the RDT accurately identified individuals with infectious syphilis under real-world conditions in an intended-use setting at the point of care. Implementing the RDT can eliminate treatment delays and may enhance disease control.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cmi.2023.06.013DOI Listing

Publication Analysis

Top Keywords

multisite prospective
8
prospective field
8
field diagnostic
8
diagnostic accuracy
8
intended-use setting
8
blood
8
rpr titres
8
titres ≥18
8
sensitivity increased
8
serum 92%
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!