Objectives: Human thyroglobulin (Tg) is widely used as a tumor marker for recurrence and metastasis of differentiated thyroid cancer (DTC). Currently, serum Tg values are measured using second-generation sandwich immunoassays (2nd-IMA). However, interference by endogenous autoantibodies to thyroglobulin (TgAbs) can lead to false-negative results or falsely low Tg values. Here, we describe a new Tg assay using the immunoassay for total antigen including complex via pretreatment (iTACT) method to prevent TgAb interference and compare it with 2nd-IMA.
Methods: Tg values were evaluated by three assays: iTACT Tg, Elecsys Tg-II, which is a 2nd-IMA, and LC-MS/MS (Liquid chromatography tandem-mass spectrometry). The ratio of Tg values between each assay was then compared to the Tg value by LC-MS/MS and TgAb titer. Tg immunoreactivity was analyzed by size-exclusion chromatography.
Results: Correlation between iTACT Tg and LC-MS/MS using TgAb-positive specimens was good: Passing-Bablok regression with iTACT Tg = 1.084 × LC-MS/MS + 0.831. Correlation between 2nd-IMA and LC-MS/MS showed a relatively lower slope: 2nd-IMA = 0.747 × LC-MS/MS - 0.518. Thus, Tg values determined by iTACT Tg are equivalent to those of LC-MS/MS regardless of TgAb titer, whereas 2nd-IMA gave lower Tg values due to TgAb interference. Tg-TgAb complexes of various molecular weights were verified by size-exclusion chromatography. Tg values measured by 2nd-IMA fluctuated depending on the molecular weight of the Tg-TgAb complexes, whereas iTACT Tg accurately quantified Tg values regardless of the size of the Tg-TgAb complexes.
Conclusion: Tg values in TgAb-positive specimens were accurately determined by iTACT Tg. TgAb-positive specimens contain Tg-TgAb complexes of various molecular weights that interfere with Tg value determination by 2nd-IMA, whereas iTACT Tg is unaffected by the presence of Tg-TgAb complexes.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.clinbiochem.2023.110598 | DOI Listing |
Clin Biochem
August 2023
Development Division, FUJIREBIO INC., Tokyo, Japan; Development Division, Advanced Life Science Institute, Inc., Tokyo, Japan. Electronic address:
Objectives: Human thyroglobulin (Tg) is widely used as a tumor marker for recurrence and metastasis of differentiated thyroid cancer (DTC). Currently, serum Tg values are measured using second-generation sandwich immunoassays (2nd-IMA). However, interference by endogenous autoantibodies to thyroglobulin (TgAbs) can lead to false-negative results or falsely low Tg values.
View Article and Find Full Text PDFJ Appl Lab Med
November 2021
Research and Development Division, FUJIREBIO INC., Tokyo, Japan.
Background: Recently, second-generation thyroglobulin (Tg) sandwich immunoassays have been used in clinical laboratories to measure the serum Tg levels, which is a tumor marker used to monitor postoperative patients with differentiated thyroid cancers. However, these immunoassays are often subject to Tg autoantibody (TgAb) interference. TgAb interference is inevitable for almost all Tg immunoassays, resulting in unreliable Tg measurement values of TgAb-positive samples.
View Article and Find Full Text PDFHuman thyroglobulin (Tg) immune complexes (TgIC) were preformed at different ratios from purified Tg, and Tg antibodies (TgAb) obtained from six patients with autoimmune thyroid diseases. TgIC were characterized by precipitation with polyethylene glycol, and sedimentation in sucrose gradient. TgIC from one patient were further characterized by concanavalin A (con A), and binding to the complement factor Clq and rheumatoid factor (RF).
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!