Cardiac output is essential for patient management in critically ill patients. The state-of-the-art for cardiac output monitoring bears limitations that pertain to the invasive nature of the method, high costs, and associated complications. Hence, the determination of cardiac output in a non-invasive, accurate, and reliable way remains an unmet need. The advent of wearable technologies has directed research towards the exploitation of wearable-sensed data to improve hemodynamical monitoring. We developed an artificial neural networks (ANN)-enabled modelling approach to estimate cardiac output from radial blood pressure waveform. data including a variety of arterial pulse waves and cardiovascular parameters from 3,818 virtual subjects were used for the analysis. Of particular interest was to investigate whether the uncalibrated, namely, normalized between 0 and 1, radial blood pressure waveform contains sufficient information to derive cardiac output accurately in an population. Specifically, a training/testing pipeline was adopted for the development of two artificial neural networks models using as input: the calibrated radial blood pressure waveform (ANN), or the uncalibrated radial blood pressure waveform (ANN). Artificial neural networks models provided precise cardiac output estimations across the extensive range of cardiovascular profiles, with accuracy being higher for the ANN. Pearson's correlation coefficient and limits of agreement were found to be equal to [0.98 and (-0.44, 0.53) L/min] and [0.95 and (-0.84, 0.73) L/min] for ANN and ANN, respectively. The method's sensitivity to major cardiovascular parameters, such as heart rate, aortic blood pressure, and total arterial compliance was evaluated. The study findings indicate that the uncalibrated radial blood pressure waveform provides sample information for accurately deriving cardiac output in an population of virtual subjects. Validation of our results using human data will verify the clinical utility of the proposed model, while it will enable research applications for the integration of the model in wearable sensing systems, such as smartwatches or other consumer devices.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10262040PMC
http://dx.doi.org/10.3389/fbioe.2023.1199726DOI Listing

Publication Analysis

Top Keywords

cardiac output
32
blood pressure
28
radial blood
24
pressure waveform
24
uncalibrated radial
12
artificial neural
12
neural networks
12
cardiac
8
cardiovascular parameters
8
virtual subjects
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!