This work presents the optimized parameters of 3D printing for print bolus using BolusCM material. Printing parameters were selected of the homogeneity and absence of air gaps. The dosimetric features of printed bolus were measured with a plane-parallel ionization chamber and EBT3 radiochromic film. Measured features were compared with those estimated with Monte Carlo methods. BolusCM shows good characteristics to be used as bolus material in radiotherapy with electrons, where the printing process allows personalizing the bolus in function of patient characteristics. The material low-cost, the 3D printing and the dosimetric features are few of the advantages of using BolusCM material in radiotherapy with electrons in skin cancer treatment.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.apradiso.2023.110899 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!