Objectives: The aim of the present study was to evaluate the accuracy of a non-contact semiconductor X-ray analyzer for quality assurance in intraoral radiography, especially a comparison with an ionization chamber dosimeter.
Methods: Intraoral radiography was performed with intraoral X-ray unit using the dental protocol at our hospital: tube voltage, 70 kV; tube current, 7 mA. Accuracy of dose and half-value layer (HVL) measurements was analyzed with a non-contact semiconductor X-ray analyzer and an ionization chamber dosimeter. Stability of the semiconductor sensor, effect of scattered radiation, and comparison of measured HVL between the ionization chamber and the semiconductor sensor were analyzed in this study.
Results: The values with the semiconductor sensor were tube voltage: 70.3 ± 0.2 kVp (degree of variability: 0.28%), dose: 454.1 ± 12.3 μGy (degree of variability: 2.7%), and HVL: 1.91 ± 0.02 mmAl (degree of variability: 1.0%). With collimator, the dose with the semiconductor sensor and the ionization chamber decreased by 2.3 μ Gy and 5.2 μ Gy, respectively. The measured HVL of the semiconductor dosimeter was more than that of ionization chamber, and the semiconductor dosimeter was less than ionization chamber in variation of between without and with collimator.
Conclusion: This study indicated the accuracy of a non-contact semiconductor X-ray analyzer for quality assurance in intraoral radiography, especially a comparison with an ionization chamber dosimeter. The semiconductor sensor can be useful for quality assurance in intraoral radiography.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s11282-023-00692-9 | DOI Listing |
Int J Part Ther
March 2025
Institute of Medical Physics and Radiation Protection, University of Applied Sciences, Giessen, Germany.
Purpose: The spot size of scanned particle beams is of crucial importance for the correct dose delivery and, therefore, plays a significant role in the quality assurance (QA) of pencil beam scanning ion beam therapy.
Materials And Methods: This study compares 5 detector types-radiochromic film, ionization chamber (IC) array, flat panel detector, multiwire chamber, and IC-for measuring the spot size of proton and carbon ion beams.
Results: Variations of up to 30% were found between detectors, underscoring the impact of detector choice on QA outcomes.
Tech Innov Patient Support Radiat Oncol
March 2025
Graduate School of Biomedical Sciences, Tokushima University, Tokushima, Tokushima 770-8503, Japan.
Purpose: This study aims to compare treatment plans created using RapidPlan and PlanIQ for twelve patients with prostate cancer, focusing on dose uniformity, dose reduction to organs at risk (OARs), plan complexity, and dose verification accuracy. The goal is to identify the tool that demonstrates superior performance in achieving uniform target dose distribution and reducing OAR dose, while ensuring accurate dose verification.
Methods: Dose uniformity in the planning target volume, excluding the rectum, and dose reduction in the OARs (the rectum and bladder) were assessed.
J Neurosci Methods
January 2025
Cognitive Neuroscience Laboratory, German Primate Center - Leibniz Institute for Primate Research, Goettingen, Germany; Faculty of Biology and Psychology, University of Goettingen, Goettingen 37077 Germany.
Background: Neuronal activity is modulated by behavior and cognitive processes. The combination of several neurotransmitter systems, acting directly or indirectly on specific populations of neurons, underlie such modulations. Most studies with non-human primates (NHPs) fail to capture this complexity, partly due to the lack of adequate methods for reliably and simultaneously measuring a broad spectrum of neurotransmitters while the animal engages in behavioral tasks.
View Article and Find Full Text PDFIndian J Nucl Med
November 2024
Center for Research and Production of Radioisotopes, Dalat Nuclear Research Institute, Vietnam Atomic Energy Institute (VINATOM), Da Lat City, Lam Dong Province, Vietnam.
Aims: This study aimed to evaluate the principal technical characteristics of a well-type gas-filled ionization chamber dose calibrator used in measuring radiopharmaceutical activity, namely accuracy, repeatability, and linearity. Furthermore, this work also explored the correlation between the device's response and the position and volume of the radiopharmaceutical I-131.
Materials And Methods: Experimental measurements were conducted on the ATOMLAB 500 dose calibrator using NIST traceable Cs-137 source to determine the accuracy and repeatability.
Anal Chim Acta
January 2025
University Regensburg, Institute of Analytical Chemistry, Universitätsstrasse 31, 93053, Regensburg, Germany. Electronic address:
Background: The demand for lithium-ion cells in the automotive industry is rapidly growing due to the increasing electrification of the transportation sector. The electrolyte composition plays a critical role in determining the lifetime and performance of these large-format cells. Additionally, advancements in this field are leading to frequent changes in both electrode materials and electrolyte formulations.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!