Objectives: To analyze device designs, similarities and overlaps of custom-made fenestrated arch endografts intended for mid/distal arch thoracic endovascular aortic repair.

Materials And Methods: A multicenter cross-sectional study analyzing custom-made anonymized graft plans was performed. Graft plans were included from a cohort of mid/distal aortic arch repairs using custom-made fenestrated aortic endografts treated at 8 centers. Grafts targeting >2 arteries were excluded. No patient/clinical data were analyzed. A descriptive analysis was performed followed by an analysis of overlap of the designs to reach a common design in which the greatest number of grafts would overlap.

Results: One hundred thirty-one graft plans were included. All grafts were custom-made grafts from the COOK Medical Fenestrated arch platform. Ninety-four (71.8%) had a scallop-and-single-fenestration design, 33 (25.2%) had a single fenestration and 4 (4.3%) a single scallop. For analysis purposes, these latter 4 grafts were excluded. Two main graft plans () were proposed after analysis with similar configuration (1 scallop with 30 mm width, 20 mm height, 12:00 position; 1 preloaded fenestration with 8 mm diameter, 26 mm from the top of the graft and 12:00 position; tapered, with a 193 mm length and 32 mm distal diameter) but with 2 different proximal diameters of 38 mm () and 44 mm (), reaching an overall feasibility of 85.8% (n=109), being 47.2% (n=60) and 38.6% (n=49) for each design, respectively.

Conclusion: The degree of overlap between the studied fenestrated and/or scalloped thoracic endovascular aneurysm repair (TEVAR) graft designs was high. Future studies analyzing these designs in a real-world cohort of patients are needed to further address off-the-shelf feasibility.

Clinical Impact: In a multicenter study analyzing 127 fenestrated aortic arch endograft plans from 9 aortic centers, we found that the degree of overlap between the studied fenestrated and/or scalloped arch graft designs was high and that 2 proposed graft designs would be theoretically applicable in 85.8% of cases. Future studies analyzing these designs in a real-world cohort of patients are needed to further address off-the-shelf feasibility.

Download full-text PDF

Source
http://dx.doi.org/10.1177/15266028231179593DOI Listing

Publication Analysis

Top Keywords

graft plans
16
custom-made fenestrated
12
fenestrated arch
12
graft designs
12
thoracic endovascular
8
study analyzing
8
graft
8
plans included
8
aortic arch
8
fenestrated aortic
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!