Background: Single-use flexible bronchoscopes(SFB) eliminate the risk of bronchoscopy-related infection compared with traditional reusable flexible bronchoscopes(RFB). At present, there is no comparative study between SFB and RFB in the aspects of biopsy and interventional therapy. This study aims to explore whether SFB can perform complex bronchoscopic procedures such as transbronchial biopsies just like RFB.

Methods: We conducted a prospective controlled study. A total of 45 patients who required bronchoscopic biopsy in our hospital from June 2022 to December 2022 were enrolled. The patients were divided into the SFB group and the RFB group, and routine bronchoscopy, bronchoalveolar lavage, and biopsy were performed respectively. Data on the time of routine bronchoscopy, the recovery rate of bronchoalveolar lavage fluid(BALF), biopsy time, and bleeding volume were collected. Then we used the two-sample t-test and the χ test to assess the performance differences between SFB and RFB. We also designed a questionnaire to compare the performance between SFB and RFB by different bronchoscope operators.

Results: The routine examination time of SFB and RFB was 3.40 ± 0.50 min and 3.55 ± 0.42 min, respectively. There was no significant difference between the two groups (P = 0.308). The recovery rate of BALF was (46.56 ± 8.22) % in the SFB group and (47.00 ± 8.07) in the RFB group, without a significant difference between the two groups(P = 0.863). The biopsy time was similar(4.67 ± 0.51 min VS 4.57 ± 0.45 min) in both groups, with no significant difference(P = 0.512). The positive biopsy rate was 100% in both groups, with no significant difference. Overall, the bronchoscope operators were generally satisfied with SFB.

Conclusion: SFBs are non-inferior to RFBs in routine bronchoscopy, bronchoalveolar lavage, and biopsy. It is suggested that SFBs have a wider clinical application.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10252171PMC
http://dx.doi.org/10.1186/s12890-023-02478-5DOI Listing

Publication Analysis

Top Keywords

sfb rfb
16
routine bronchoscopy
12
bronchoalveolar lavage
12
single-use flexible
8
flexible bronchoscopes
8
traditional reusable
8
reusable flexible
8
prospective controlled
8
controlled study
8
sfb group
8

Similar Publications

Background: Single-use flexible bronchoscopes(SFB) eliminate the risk of bronchoscopy-related infection compared with traditional reusable flexible bronchoscopes(RFB). At present, there is no comparative study between SFB and RFB in the aspects of biopsy and interventional therapy. This study aims to explore whether SFB can perform complex bronchoscopic procedures such as transbronchial biopsies just like RFB.

View Article and Find Full Text PDF

Background And Objective: Until 2009, only reusable bronchoscopes were marketed, but the introduction and widespread adoption of single-use flexible bronchoscopes (SFBs) as an emerging technology has since accelerated. Several studies have described the costs of reusable flexible bronchoscopes (RFBs) and SFBs. This meta-analysis aimed to compile the current published evidence to analyse the cost of different scenarios using RFBs and SFBs.

View Article and Find Full Text PDF

Faba bean flours (germinated, soaked, cooked and raw) were evaluated for physiochemical and functional properties. The flour samples exhibited considerable amounts of carbohydrates (58.79-66.

View Article and Find Full Text PDF

Design Principles and Developments of Integrated Solar Flow Batteries.

Acc Chem Res

November 2020

Department of Chemistry, University of Wisconsin-Madison, 1101 University Avenue, Madison, Wisconsin 53706, United States.

ConspectusDue to the intermittent nature of sunlight, practical round-trip solar energy utilization systems require both efficient solar energy conversion and inexpensive large-scale energy storage. Conventional round-trip solar energy utilization systems typically rely on the combination of two or more separated devices to fulfill such requirements. Integrated solar flow batteries (SFBs) are a new type of device that integrates solar energy conversion and electrochemical storage.

View Article and Find Full Text PDF

Production of GFP and glucoamylase by recombinant Aspergillus niger: effects of fermentation conditions on fungal morphology and protein secretion.

Biotechnol Prog

February 2006

Department of Chemical and Biomolecular Engineering, The Ohio State University, 140 West 19th Avenue, Columbus, Ohio 43210, USA.

The effect of filamentous fungal morphology on heterologous protein secretion was investigated using the recombinant Aspergillus niger strain AB4.1[pgpdAGLAGFP], which contained the gene coded for the GLA-GFP (glucoamylase-green fluorescence protein) fusion protein. Three culturing systems were studied to develop different morphological forms of the fungus.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!