Bioprosthetic Valve Remodeling in Nonfracturable Surgical Valves: Impact on THV Expansion and Hydrodynamic Performance.

JACC Cardiovasc Interv

Centre for Cardiovascular Innovation, St. Paul's and Vancouver General Hospital, Vancouver, Canada; Cardiovascular Translational Laboratory, Providence Research and Centre for Heart Lung Innovation, Vancouver, Canada; Centre for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada. Electronic address:

Published: July 2023

Background: There are limited data on the effect of bioprosthetic valve remodeling (BVR) on transcatheter heart valve (THV) expansion and function following valve-in-valve (VIV) transcatheter aortic valve replacement (TAVR) in a nonfracturable surgical heart valve (SHV).

Objectives: This study sought to assess the impact of BVR of nonfracturable SHVs on THVs after VIV implantation.

Methods: VIV TAVR was performed using 23-mm SAPIEN3 (S3, Edwards Lifesciences) or 23/26-mm Evolut Pro (Medtronic) THVs implanted in 21/23-mm Trifecta (Abbott Structural Heart) and 21/23-mm Hancock (Medtronic) SHVs with BVR performed with a noncompliant TRUE balloon (Bard Peripheral Vascular Inc). Hydrodynamic assessment was performed, and multimodality imaging including micro-computed tomography was performed before and after BVR to assess THV and SHV expansion.

Results: BVR resulted in limited improvement of THV expansion. The largest gain in expansion was observed for the S3 in the 21-mm Trifecta with up to a 12.7% increase in expansion at the outflow of the valve. Minimal change was observed at the level of the sewing ring. The Hancock was less amenable to BVR with lower final expansion dimensions than the Trifecta. BVR also resulted in notable surgical post flaring of up to 17.6°, which was generally more marked with the S3 than with the Evolut Pro. Finally, BVR resulted in very limited improvement in hydrodynamic function. Severe pinwheeling was observed with the S3, which improved slightly but persisted despite BVR.

Conclusions: When performing VIV TAVR inside a Trifecta and Hancock SHV, BVR had a limited impact on THV expansion and resulted in SHV post flaring with unknown consequences on coronary obstruction risk and long-term THV function.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcin.2023.03.035DOI Listing

Publication Analysis

Top Keywords

thv expansion
16
bvr limited
12
bvr
9
bioprosthetic valve
8
valve remodeling
8
impact thv
8
heart valve
8
viv tavr
8
evolut pro
8
limited improvement
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!