Purpose: To investigate the possible risk factors for treatment failure in patients who had undergone Preserflo Microshunt (PMS) implantation, using anterior-segment optical coherence tomography (AS-OCT) to analyze the internal structures of the bleb.
Methods: The PMS blebs of 54 patients were evaluated with AS-OCT. A mathematical model was used to calculate the total filtering surface of the episcleral fluid cavity (EFC) and the hydraulic conductivity (HC) of the bleb wall. Complete and qualified success were defined as IOP between 6 and 17 mmHg with or without glaucoma medication. The relation between baseline characteristics and probability of bleb success was analyzed by bivariate and multivariate logistic regression. The main outcome measures were mean bleb wall thickness (BWT), reflectivity (BWR), HC, mean horizontal and vertical diameter and total filtering surface (TFS) of the EFC.
Results: Blebs from 74% patients were considered as complete success and 26% as failure. BWR and BWT increased linearly up to the first year in both groups. BWR was higher in the group failure (p = 0.02) and BWT in the group success (p<0.001). EFC was wider and shorter in the success group (p = 0.009, p = 0.03). Higher TFS showed a negative correlation with IOP (r = -0.4, p = 0.002). Higher baseline IOP was associated with success of PMS by multivariate analysis (p = 0.01). Mean HC, 0.034 ± 0.008 (μL/min)/mm2/mmHg, was negatively correlated with bleb surface (r = -0.5, p<0.0001) and wall´s thickness (r = -0.3, p = 0.01).
Conclusions: AS-OCT revealed that successful PMS blebs could show either thick hyporreflective walls or wide filtering surfaces with thin capsules. A higher baseline IOP increased the probability of surgical success.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10249890 | PMC |
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0286884 | PLOS |
Jpn J Ophthalmol
January 2025
Department of Ophthalmology, Graduate of Medicine, Faculty of Medicine, The University of Tokyo, Tokyo, Japan.
Purpose: To assess the efficacy and safety of PreserFlo MicroShunt (PMS) combined with mitomycin C in patients with medically treated primary open-angle glaucoma (POAG).
Study Design: A retrospective observational study.
Methods: The study examined 83 eyes from 83 patients with medically treated glaucoma surgery naive POAG.
BMC Ophthalmol
January 2025
Department of Ophthalmology, Medical Faculty, University Hospital of Cologne, Kerpener Strasse 62, 50937, Cologne, Germany.
Background/ Aims: To analyze the longitudinal change in Bruch's membrane opening minimal rim width (BMO-MRW) and peripapillary retinal nerve fiber layer (pRNFL) thickness using optical coherence tomography (OCT) after implantation of a PRESERFLO® microshunt for surgical glaucoma management in adult glaucoma patients.
Methods: Retrospective data analysis of 59 eyes of 59 participants undergoing implantation of a PRESERFLO microshunt between 2019 and 2022 at a tertiary center for glaucoma management. Surgical management included primary temporary occlusion of the glaucoma shunt to prevent early hypotony.
Sci Rep
January 2025
Department of Ophthalmology and ORL, Faculty of Medicine, Clinico San Carlos Hospital, Complutense University, Clinico San Carlos Hospital Health Research Institute (IdISSC), Calle del Prof Martín Lagos, s/n,, Madrid, 28040, Spain.
This study compared the efficacy safety profiles of the Xen 63 and Preserflo MicroShunt devices, both standalone, in patients with primary open-angle glaucoma (POAG). It is a retrospective and single-center study conducted on consecutive on patients with medically uncontrolled POAG who underwent either a standalone Xen 63 or a standalone Preserflo and had a 12-month follow-up visit. The primary outcome was the mean IOP at month-12.
View Article and Find Full Text PDFCureus
November 2024
Department of Ophthalmology and Visual Science, Chiba University Graduate School of Medicine, Chiba, JPN.
Preserflo MicroShunt (PMS) implantation is a minimally invasive surgical procedure for treating glaucoma. Postoperative hypotony, a common complication of PMS implantation, can be prevented and treated with 10-0 nylon insertion. In this report, we present a case of postoperative hypotony following PMS implantation that was treated with intraluminal insertion of 9-0 nylon.
View Article and Find Full Text PDFOcul Immunol Inflamm
December 2024
Centre for Ophthalmology, University Hospital Tuebingen, Tuebingen, Germany.
Purpose: To investigate the efficacy and safety of the Preserflo MicroShunt (PMS) in treating glaucoma secondary to different uveitic glaucoma (UG) types in a European population.
Methods: This study evaluated consecutive eyes with UG that received the PMS. The primary outcome measure was 12-month surgical success (≥20% IOP reduction, target IOP of 6-21 mmHg).