Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objective: To describe the development, implementation, and validation of a radiology-administered protocol to obtain magnetic resonance imaging (MRI) in patients with cochlear implants and auditory brainstem implants without magnet removal.
Study Design: Retrospective review and description of novel care pathway.
Methods: A radiology-administered protocol was designed based on careful input from the radiology safety committee and neurotology. Radiology technologist training modules, consent instructions, patient educational material, clinical audits, and other safeguards were implemented, with samples provided in this report. The primary outcomes measured included instances of magnet displacement during MRI and premature termination of MRI studies secondary to pain.
Results: Between June 19, 2018, and October 12, 2021, 301 implanted ears underwent MRI without magnet removal, including 153 devices housing diametric MRI-conditional magnets, and 148 implants with conventional axial (i.e., nondiametric) magnets. Among cases with diametric MRI-conditional magnets, all studies were completed without magnet dislodgement or need to terminate imaging early due to pain. Among cases with conventional axial (nondiametric) magnets, 29 (19.6%) MRI studies were stopped prematurely secondary to pain or discomfort; the overall rate of this event was 9.6% (29 of 301) among the entire study cohort. In addition, 6.1% (9 of 148) experienced confirmed magnet displacement despite headwrap placement; the overall rate among all cases was 3.0% (9 of 301). Eight of these patients received successful external magnet reseating through manual pressure on the external scalp without surgery, and one required surgical replacement of the magnet in the operating room. There were no documented instances of hematoma, infection, device or magnet extrusion, internal device movement (i.e., gross receiver-stimulator migration), or device malfunction in this cohort related to MRI.
Conclusions: We present the successful implementation of a radiology-administered protocol designed to streamline care for cochlear implant and auditory brainstem implant recipients who require MRI and ease clinical demands for otolaryngology providers. Examples of resources developed, including a process map, radiology training modules, consent instructions, patient educational materials, clinical audit, and other procedural safety measures are provided so interested groups may consider adapting and implementing related measures according to need.