Five-year safety and efficacy of leadless pacemakers in a Dutch cohort.

Heart Rhythm

Department of Cardiology, Amsterdam UMC Location AMC, Amsterdam; The Netherlands, Amsterdam Cardiovascular Sciences, Heart Failure & Arrhythmias, Amsterdam, The Netherlands.

Published: August 2023

Background: Adequate real-world safety and efficacy of leadless pacemakers (LPs) have been demonstrated up to 3 years after implantation. Longer-term data are warranted to assess the net clinical benefit of leadless pacing.

Objective: The purpose of this study was to evaluate the long-term safety and efficacy of LP therapy in a real-world cohort.

Methods: In this retrospective cohort study, all consecutive patients with a first LP implantation from December 21, 2012, to December 13, 2016, in 6 Dutch high-volume centers were included. The primary safety endpoint was the rate of major procedure- or device-related complications (ie, requiring surgery) at 5-year follow-up. Analyses were performed with and without Nanostim battery advisory-related complications. The primary efficacy endpoint was the percentage of patients with a pacing capture threshold ≤2.0 V at implantation and without ≥1.5-V increase at the last follow-up visit.

Results: A total of 179 patients were included (mean age 79 ± 9 years), 93 (52%) with a Nanostim and 86 (48%) with a Micra VR LP. Mean follow-up duration was 44 ± 26 months. Forty-one major complications occurred, of which 7 were not advisory related. The 5-year major complication rate was 4% without advisory-related complications and 27% including advisory-related complications. No advisory-related major complications occurred a median 10 days (range 0-88 days) postimplantation. The pacing capture threshold was low in 163 of 167 patients (98%) and stable in 157 of 160 (98%).

Conclusion: The long-term major complication rate without advisory-related complications was low with LPs. No complications occurred after the acute phase and no infections occurred, which may be a specific benefit of LPs. The performance was adequate with a stable pacing capture threshold.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hrthm.2023.05.031DOI Listing

Publication Analysis

Top Keywords

advisory-related complications
16
safety efficacy
12
pacing capture
12
capture threshold
12
complications occurred
12
efficacy leadless
8
leadless pacemakers
8
complications
8
major complications
8
major complication
8