Background: Early cancer detection is crucial for patients' survival. The image quality in In-octreotide SPECT imaging could be improved by using Monte Carlo (MC)-based reconstruction. The aim of this observational study was to determine the detection rate of simulated liver lesions for MC-based ordered subset expectation maximization (OSEM) reconstruction compared to conventional attenuation-corrected OSEM reconstruction.
Methods: Thirty-seven SPECT/CT examinations with In-octreotide were randomly selected. The inclusion criterion was no liver lesions at the time of examination and for the following 3 years. SPECT images of spheres representing lesions were simulated using MC. The raw data of the spheres were added to the raw data of the established healthy patients in 26 of the examinations, and the remaining 11 examinations were not modified. The images were reconstructed using conventional OSEM reconstruction with attenuation correction and post filtering (fAC OSEM) and MC-based OSEM reconstruction without and with post filtering (MC OSEM and fMC OSEM, respectively). The images were visually and blindly evaluated by a nuclear medicine specialist. The criteria evaluated were liver lesion yes or no, including coordinates if yes, with confidence level 1-3. The percentage of detected lesions and accuracy (percentage of correctly classified cases), as well as tumor-to-normal tissue concentration (TNC) ratios and signal-to-noise ratios (SNRs), were evaluated.
Results: The detection rates were 30.8% for fAC OSEM, 42.3% for fMC OSEM, and 50.0% for MC OSEM. The accuracies were 45.9% for fAC OSEM, 45.9% for fMC OSEM, and 54.1% for MC OSEM. The number of false positives was higher for fMC and MC OSEM. The observer's confidence level was higher in filtered images than in unfiltered images. TNC ratios were significantly higher, statistically, with MC OSEM and fMC OSEM than with AC OSEM, but SNRs were similar due to higher noise with MC OSEM.
Conclusion: One in two lesions were found using MC OSEM versus one in three using conventional reconstruction. TNC ratios were significantly improved, statistically, using MC-based reconstruction, but the noise levels increased and consequently the confidence level of the observer decreased. For further improvements, image noise needs to be suppressed.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10238363 | PMC |
http://dx.doi.org/10.1186/s40658-023-00557-4 | DOI Listing |
F-Florbetaben (FBB) uptake in the supratentorial cortex is indicative of amyloid positivity. Due to PET's low spatial resolution, image noise, and spill-over of signal from adjacent white-matter into gray-matter, there are inconsistencies in ratings among trained readers. A set of 264 F-Florbetaben (amyloid) PET/MRI exams were reconstructed using conventional ordered subset expectation maximization (OSEM) method and MR-guided block sequential regularized expectation maximization (MRgBSREM) method.
View Article and Find Full Text PDFAsia Ocean J Nucl Med Biol
January 2025
Department of Radiology, Faculty of Medicine, Shimane University, Izumo, Japan.
Objectives: We investigated image quality and standardized uptake values (SUVs) for different lesion sizes using clinical data generated by F-FDG-prone breast silicon photomultiplier (SiPM)-based positron emission tomography/computed tomography (PET/CT).
Methods: We evaluated the effect of point-spread function (PSF) modeling and Gaussian filtering (Gau) and determined the optimal reconstruction conditions. We compared the signal-to-noise ratio (SNR), contrast, %coefficient of variation (%CV), SUV, and Likert scale score between ordered-subset expectation maximization (OSEM) time-of-flight (TOF) and OSEM+TOF+PSF in phantom and clinical studies.
Clin Physiol Funct Imaging
January 2025
Department of Translational Medicine, Wallenberg Center for Molecular Medicine, Lund University, Malmö, Sweden.
Background: Positron emission tomography (PET) response criteria in solid tumours (PERCIST 1.0) is a systematic assessment of therapy response with fluorine 18 fluorodeoxyglucose ([F]FDG) PET. The ordered subset expectation maximization (OSEM) and block sequential regularized expectation maximization (BSREM) reconstruction algorithms are both commonly used.
View Article and Find Full Text PDFAnn Nucl Med
December 2024
Division of Functional Imaging, Exploratory Oncology Research and Clinical Trial Center, National Cancer Center, 6-5-1 Kashiwanoha, Kashiwa, 277-8577, Japan.
Objective: To investigate the clinical utility of a new anthropomorphic phantom that reproduces the chest and abdomen better than the conventional National Electrical Manufacturers Association (NEMA) body phantom, count rates and image quality of PET images obtained from patients were evaluated.
Methods: Anthropomorphic phantoms were used to include radioactivity in the lung, liver, kidney, and background regions. Two NEMA body phantoms were used for chest and abdominal assessments.
Med Phys
November 2024
High-Field and Hybrid MR Imaging, University Hospital Essen, University of Duisburg-Essen, Essen, Germany.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!