Anemia and hypoxemia are common clinical conditions that are difficult to study and may impact pulse oximeter performance. Utilizing an in vitro circulation system, we studied performance of three pulse oximeters during hypoxemia and severe anemia. Three oximeters including one benchtop, one handheld, and one fingertip device were selected to reflect a range of cost and device types. Human blood was diluted to generate four hematocrit levels (40%, 30%, 20%, and 10%). Oxygen and nitrogen were bubbled through the blood to generate a range of oxygen saturations (OHb) and the blood was cycled through the in vitro circulation system. Pulse oximeter saturations (SpO) were paired with simultaneously-measured OHb readings from a reference CO-oximeter. Data for each hematocrit level and each device were least-squares fit to a 2nd-order equation with quality of each curve fit evaluated using standard error of the estimate. Bias and average root mean square error were calculated after correcting for the calibration difference between human and in vitro circulation system calibration. The benchtop oximeter maintained good accuracy at all but the most extreme level of anemia. The handheld device was not as accurate as the benchtop, and inaccuracies increased at lower hematocrit levels. The fingertip device was the least accurate of the three oximeters. Pulse oximeter performance is impacted by severe anemia in vitro. The use of in vitro calibration systems may play an important role in augmenting in vivo performance studies evaluating pulse oximeter performance in challenging conditions.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10651532PMC
http://dx.doi.org/10.1007/s10877-023-01031-3DOI Listing

Publication Analysis

Top Keywords

pulse oximeter
20
vitro circulation
16
circulation system
16
severe anemia
12
oximeter performance
12
hypoxemia severe
8
anemia vitro
8
three oximeters
8
fingertip device
8
hematocrit levels
8

Similar Publications

Community Health Workers (CHWs) in low- and middle-income countries are essential in providing primary health care to remote communities. However, due to limited diagnostic tools, CHWs often struggle to correctly identify childhood illnesses, especially pneumonia. We conducted a prospective pilot study and used qualitative research methods to evaluate acceptability and feasibility of a multimodal pulse oximeter used by CHWs during their integrated community case management (iCCM) of childhood illness consultations in rural Burundi.

View Article and Find Full Text PDF

Annual surveys of refugees in Gambella, Ethiopia suggest that anemia is a persistent public health problem among non-pregnant women of reproductive age (NP-WRA, 15-49 years). Measurement of anemia in most refugee camp settings is conducted using an invasive HemoCue 301. We assessed the accuracy and precision of a non-invasive, pulse CO-oximeter in measuring anemia among NP-WRA in four Gambella refugee camps.

View Article and Find Full Text PDF

Background: Audio-analgesia is one of the unexplored aspects of behavior management in pediatric dentistry, and recently, there have been many new inclusions in shades of noise that were previously just limited to white noise, like brown noise and pink noise.

Aim: The aim of the study was to evaluate and compare the effectiveness of white noise, brown noise, and pink noise on dental anxiety in pediatric patients undergoing primary tooth extraction.

Study Settings And Design: Multiarm, triple-blinded, parallel-group randomized controlled trial.

View Article and Find Full Text PDF

Validation of Aerobic Capacity (VO2max) and Pulse Oximetry in Wearable Technology.

Sensors (Basel)

January 2025

School of Integrated Health Sciences, Department of Kinesiology and Nutrition Sciences, University of Nevada, Las Vegas, NV 89154, USA.

Introduction: As wearable technology becomes increasingly popular and sophisticated, independent validation is needed to determine its accuracy and potential applications. Therefore, the purpose of this study was to evaluate the accuracy (validity) of VO2max estimates and blood oxygen saturation measured via pulse oximetry using the Garmin fēnix 6 with a general population participant pool.

Methods: We recruited apparently healthy individuals (both active and sedentary) for VO2max (n = 19) and pulse oximetry testing (n = 22).

View Article and Find Full Text PDF

Accuracy of Rhythm Diagnostic Systems' MultiSense in Detection of Arterial Oxygen Saturation and Respiratory Rate During Hypoxia in Humans: Effects of Skin Color and Device Localization.

Sensors (Basel)

December 2024

Biomedicine Research Center of Strasbourg (CRBS), UR 3072, "Mitochondria, Oxidative Stress and Muscle Plasticity", Faculty of Medicine, University of Strasbourg, 67000 Strasbourg, France.

The continuous monitoring of oxygen saturation (SpO) and respiratory rates (RRs) are major clinical issues in many cardio-respiratory diseases and have been of tremendous importance during the COVID-19 pandemic. The early detection of hypoxemia was crucial since it precedes significant complications, and SpO follow-up allowed early hospital discharge in patients needing oxygen therapy. Nevertheless, fingertip devices showed some practical limitations.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!