Swept-source optical coherence tomography (SS-OCT) demonstrates superior performance in comparison to spectral domain OCT with regard to depth ranging. The main driver of cost for SS-OCT systems is, however, the price of the source. Here we show a low-cost alternative swept source that uses a thermally tuned vertical-cavity surface-emitting laser (VCSEL) at 850 nm. Its center wavelength can be tuned by adjusting the operating temperature through modulation of the injection current. At 2 kHz sweep rate, the depth range of the system was 5 cm, with a sensitivity roll-off of under -3 dB across this range. The system achieved a sensitivity of 97 dB with a sample beam power of 0.3 mW and an axial resolution of 50 µm in air. To demonstrate the system performance in vivo, an eye of a healthy volunteer was measured, and full-eye scans were acquired at 25 and 50 kHz from the cornea to the retina. Based on our results, we believe that this technology can be used as a cost-effective alternative OCT for point-of-care diagnostics.

Download full-text PDF

Source
http://dx.doi.org/10.1364/OL.489050DOI Listing

Publication Analysis

Top Keywords

thermally tuned
8
vcsel 850 nm
8
low-cost alternative
8
range system
8
tuned vcsel
4
850 nm low-cost
4
alternative source
4
source full-eye
4
full-eye ss-oct
4
ss-oct swept-source
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!