AI Article Synopsis

  • The study aims to evaluate the occurrence of radiolucent lines (RLLs) and their connection to loosening in the ATTUNE TKA after total knee arthroplasty (TKA).
  • A systematic search of 263 studies identified 12 relevant ones, involving 3,861 TKAs, revealing a 21.4% overall rate of RLLs, with higher rates in specific designs, such as 27.4% for cruciate-retaining and 29.9% for fixed-bearing types.
  • Although the ATTUNE TKA showed higher rates of RLLs compared to other implants, the actual rates of loosening and revision are low (1.2% and 0.9

Article Abstract

Aims: Loosening of components after total knee arthroplasty (TKA) can be associated with the development of radiolucent lines (RLLs). The aim of this study was to assess the rate of formation of RLLs in the cemented original design of the ATTUNE TKA and their relationship to loosening.

Methods: A systematic search was undertaken using the Cochrane methodology in three online databases: MEDLINE, Embase, and CINAHL. Studies were screened against predetermined criteria, and data were extracted. Available National Joint Registries in the Network of Orthopaedic Registries of Europe were also screened. A random effects model meta-analysis was undertaken.

Results: Of 263 studies, 12 were included with a total of 3,861 TKAs. Meta-analysis of ten studies showed high rates of overall tibial or femoral RLLs for the cemented original design of the ATTUNE TKA. The overall rate was 21.4% (95% confidence interval (CI) 12.7% to 33.7%) for all types of design but was higher for certain subgroups: 27.4% (95% CI 13.4% to 47.9%) for the cruciate-retaining type, and 29.9% (95% CI 15.6% to 49.6%) for the fixed-bearing type. Meta-analysis of five studies comparing the ATTUNE TKA with other implants showed a significantly higher risk of overall tibial or femoral RLLs (odds ratio (OR) 2.841 (95% CI 1.219 to 6.623); p = 0.016) for the ATTUNE. The rates of loosening or revision for loosening were lower, at 1.2% and 0.9% respectively, but the rates varied from 0% to 16.3%. The registry data did not report specifically on the original ATTUNE TKA or on revision due to loosening, but 'all-cause' five-year revision rates for the cemented ATTUNE varied from 2.6% to 5.9%.

Conclusion: The original cemented ATTUNE TKA has high rates of RLLs, but their clinical significance is uncertain given the overall low associated rates of loosening and revision. However, in view of the high rates of RLLs and the variation in the rates of loosening and revision between studies and registries, close surveillance of patients who have undergone TKA with the original ATTUNE system is recommended.

Download full-text PDF

Source
http://dx.doi.org/10.1302/0301-620X.105B6.BJJ-2022-0675.R1DOI Listing

Publication Analysis

Top Keywords

attune tka
20
cemented original
12
original design
12
design attune
12
high rates
12
rates loosening
12
loosening revision
12
attune
9
radiolucent lines
8
total knee
8

Similar Publications

Article Synopsis
  • This study examined the clinical and radiographic results of two tibial component designs (Attune S and Attune S+) in total knee arthroplasty (TKA) over an average follow-up of 4 years.
  • No significant differences were found in clinical outcomes or radiologic parameters between the two designs, and the incidence of tibial radiolucent lines (RLL) was similar in both groups.
  • However, the subgroup analysis indicated that patients with RLL had a smaller medial proximal tibial angle and more varus alignment preoperatively compared to those without RLL.
View Article and Find Full Text PDF
Article Synopsis
  • Asymmetric tibial components in total knee arthroplasty (TKA) were evaluated against symmetric components to assess bony coverage in cases with simulated bone defects using CT scans from 45 patients.
  • The study examined three models (PCCK, NexGen LCCK, and ATTUNE) with varying thickness augmentations to simulate defects of 5, 10, and 15 mm, measuring coverage, overhang, and underhang rates.
  • Results indicated that asymmetric components (PCCK) provided better overall coverage and reduced both anterior overhang and posterior underhang compared to symmetric components, potentially leading to improved long-term outcomes for patients.
View Article and Find Full Text PDF

Aims: This multicentre retrospective observational study's aims were to investigate whether there are differences in the occurrence of radiolucent lines (RLLs) following total knee arthroplasty (TKA) between the conventional Attune baseplate and its successor, the novel Attune S+, independent from other potentially influencing factors; and whether tibial baseplate design and presence of RLLs are associated with differing risk of revision.

Methods: A total of 780 patients (39% male; median age 70.7 years (IQR 62.

View Article and Find Full Text PDF
Article Synopsis
  • The study investigates the morphological characteristics of the distal femur in hemophilic arthritis (HA), noting changes due to long-term synovitis, and compares the dimensions of femoral osteotomy surfaces required for three types of knee prostheses.
  • A total of 50 HA patients were randomly selected for a retrospective analysis, compared to equal numbers of osteoarthritis (OA) patients and healthy individuals, using digital modeling to simulate and measure femur osteotomies for total knee arthroplasties.
  • Results indicate that HA patients have significantly shorter dimensions in various femoral measurements compared to OA and healthy individuals, with smaller femur section aspect ratios, suggesting a need for tailored prosthesis solutions in HA cases.
View Article and Find Full Text PDF

Purpose: Quality of care in total knee arthroplasty (TKA) between implants was assessed using a novel composite outcome measure, early optimal recovery (EOR), to indicate ideal clinical outcomes and minimal healthcare resource utilization.

Methods: Patients that underwent primary TKA in the study group (ATTUNE® Knee System) or control group (LCS® COMPLETE Knee System) were included in this retrospective, single-center study. EOR was defined as no complications, no readmissions, no extra outpatient visits, ≤ 48 h length of hospital stay (LOS), and restored range of motion and pain perception at 3-month follow-up.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!