Aims: Loosening of components after total knee arthroplasty (TKA) can be associated with the development of radiolucent lines (RLLs). The aim of this study was to assess the rate of formation of RLLs in the cemented original design of the ATTUNE TKA and their relationship to loosening.
Methods: A systematic search was undertaken using the Cochrane methodology in three online databases: MEDLINE, Embase, and CINAHL. Studies were screened against predetermined criteria, and data were extracted. Available National Joint Registries in the Network of Orthopaedic Registries of Europe were also screened. A random effects model meta-analysis was undertaken.
Results: Of 263 studies, 12 were included with a total of 3,861 TKAs. Meta-analysis of ten studies showed high rates of overall tibial or femoral RLLs for the cemented original design of the ATTUNE TKA. The overall rate was 21.4% (95% confidence interval (CI) 12.7% to 33.7%) for all types of design but was higher for certain subgroups: 27.4% (95% CI 13.4% to 47.9%) for the cruciate-retaining type, and 29.9% (95% CI 15.6% to 49.6%) for the fixed-bearing type. Meta-analysis of five studies comparing the ATTUNE TKA with other implants showed a significantly higher risk of overall tibial or femoral RLLs (odds ratio (OR) 2.841 (95% CI 1.219 to 6.623); p = 0.016) for the ATTUNE. The rates of loosening or revision for loosening were lower, at 1.2% and 0.9% respectively, but the rates varied from 0% to 16.3%. The registry data did not report specifically on the original ATTUNE TKA or on revision due to loosening, but 'all-cause' five-year revision rates for the cemented ATTUNE varied from 2.6% to 5.9%.
Conclusion: The original cemented ATTUNE TKA has high rates of RLLs, but their clinical significance is uncertain given the overall low associated rates of loosening and revision. However, in view of the high rates of RLLs and the variation in the rates of loosening and revision between studies and registries, close surveillance of patients who have undergone TKA with the original ATTUNE system is recommended.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1302/0301-620X.105B6.BJJ-2022-0675.R1 | DOI Listing |
Knee Surg Relat Res
November 2024
Department of Orthopaedic Surgery, Center for Joint Diseases, Kyung Hee University Hospital at Gangdong, Seoul, Korea.
Arthroplasty
November 2024
Department of Orthopedic Surgery, Kyushu University, Fukuoka, 812-8582, Japan.
Bone Joint J
November 2024
Department of Orthopedics and Trauma, Medical University of Graz, Graz, Austria.
Aims: This multicentre retrospective observational study's aims were to investigate whether there are differences in the occurrence of radiolucent lines (RLLs) following total knee arthroplasty (TKA) between the conventional Attune baseplate and its successor, the novel Attune S+, independent from other potentially influencing factors; and whether tibial baseplate design and presence of RLLs are associated with differing risk of revision.
Methods: A total of 780 patients (39% male; median age 70.7 years (IQR 62.
Orthop Surg
November 2024
Department of Orthopedics, The Second Affiliated Hospital of Anhui Medical University, Hefei, China.
Musculoskelet Surg
August 2024
Department of Science and Innovation, Bergman Clinics, Gooimeer 11, 1411 DE, Naarden, The Netherlands.
Purpose: Quality of care in total knee arthroplasty (TKA) between implants was assessed using a novel composite outcome measure, early optimal recovery (EOR), to indicate ideal clinical outcomes and minimal healthcare resource utilization.
Methods: Patients that underwent primary TKA in the study group (ATTUNE® Knee System) or control group (LCS® COMPLETE Knee System) were included in this retrospective, single-center study. EOR was defined as no complications, no readmissions, no extra outpatient visits, ≤ 48 h length of hospital stay (LOS), and restored range of motion and pain perception at 3-month follow-up.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!