Background: The optimal tibial fixation of anterior cruciate ligament (ACL) reconstruction (ACLR) grafts remains controversial.
Purpose/hypothesis: The purpose of this study was to compare the biomechanical characteristics of the TensionLoc (TL) cortical fixation device with the Double Spike Plate (DSP) fixation device for ACL tibial fixation using both bone-patellar tendon-bone (BTB) and quadriceps grafts. It was hypothesized that there would be no differences in biomechanical characteristics between the fixation devices regardless of graft type.
Study Design: Controlled laboratory study.
Methods: ACLR was performed on 14 matched-pair cadaveric knee specimens-7 pairs using quadriceps grafts (n = 3 male cadaveric knee specimens; n = 4 female cadaveric knee specimens; age, 51 ± 8 years) and 7 pairs using BTB grafts (n = 3 male cadaveric knee specimens; n = 4 female cadaveric knee specimens; age, 50 ± 7 years). One side of each pair was randomized to receive DSP fixation, and the contralateral side received TL fixation. Specimens underwent cyclic ramp loading (10 cycles each at 50-100 N, 50-250 N, and 50-400 N), followed by load-to-failure testing, with the tensile force in line with the tibial tunnel. Results between the 2 fixation types were compared with a paired test.
Results: For the quadriceps graft, there were no significant differences in cyclic loading or load-to-failure characteristics between fixation types (≥ .092 for all parameters). For the BTB graft, TL fixation resulted in higher stiffness than DSP at all cyclic testing cycles except for cycle 1 during 100-N loading and had lower displacement at 250-N loading (3.4 ± 0.1 vs 5.4 ± 0.3 mm; = .045). For load to failure, TL fixation resulted in higher stiffness than DSP fixation (232 ± 3.1 vs 188.4 ± 6.4 N/mm; = .046); however, all other load-to-failure parameters were not statistically different (≥ .135 for all parameters).
Conclusion: With the quadriceps tendon graft, there were no significant differences in biomechanical characteristics between TL and DSP ACL tibial fixations; however, with BTB grafts, the TL tibial fixation demonstrated greater biomechanical integrity than the DSP tibial fixation.
Clinical Relevance: The TL fixation device may provide an alternative ACL tibial fixation option for BTB and soft tissue grafts.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10226316 | PMC |
http://dx.doi.org/10.1177/23259671231169198 | DOI Listing |
Zhongguo Gu Shang
January 2025
Beijing Luhe Hospital, Capital Medical University, Beijing 101149, China.
Tibial plateau fracture is a fracture involving the proximal articular surface of the tibia, and its injury mechanism is complex, the fracture morphology is different, and it is often accompanied by different degrees of soft tissue injury, which is difficult to diagnose and treat. In recent years, the research hotspot has focused on solving the reduction and fixation of the posterior lateral column of the tibial plateau, because it has been clinically found that the residual sagittal plane after tibial plateau fracture is insufficient reduction or loss of reduction leads to knee joint dysfunction. The posterior inclination angle of the tibial plateau is an important parameter to describe the sagittal alignment of the tibia.
View Article and Find Full Text PDFZhongguo Gu Shang
January 2025
Department of Orthopaedics, Fudan University Affiliated Zhongshan Hospital Wusong Hospital, Shanghai 200940, China.
Objective: To analyze the clinical efficacy and safety of microsurgery in patients with Gustilo ⅢC open fractures.
Methods: A total of 64 patients with Gustilo ⅢC open fractures who were admitted to the hospital bewteen September 2018 and March 2022 were included, and divided into the observation group and the control group, with 32 cases in each one. In the observation group, there were 24 males and 8 females, aged from 29 to 42 years with an average of (36.
Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi
January 2025
Department of Orthopedics, the First Affiliated Hospital of Bengbu Medical University, Bengbu Anhui, 233000, P. R. China.
Objective: To analyze the effectiveness of three internal fixation methods, namely hollow screw combined with Kirschner wire tension band, hollow screw combined with anchor nail, and modified 1/3 tubular steel plate, in the treatment of avulsion fracture of tibial tubercle (AFTT) in adolescents.
Methods: Between January 2018 and September 2023, 19 adolescent AFTT patients who met the selection criteria were admitted. According to different internal fixation methods, patients were divided into group A (8 cases, hollow screw combined with Kirschner wire tension band), group B (6 cases, hollow screw combined with anchor nail), and group C (5 cases, modified 1/3 tubular steel plate).
Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi
January 2025
Department of Sports Medicine, the First Affiliated Hospital of Kunming Medical University, Kunming Yunnan, 650032, P. R. China.
Objective: To investigate the effectiveness of modified single patellar tunnel medial patella femoral ligament (MPFL) reconstruction in the treatment of recurrent patellar dislocation.
Methods: Between January 2023 and June 2023, a total of 61 patients with recurrent patellar dislocation who underwent MPFL reconstruction with autologous semitendinosus were enrolled and divided into 2 groups using random number table method. In the patellar anchor group, 31 patients were treated with MPFL reconstruction with double medial patellar anchors, and 30 patients in the patellar tunnel group were treated with MPFL reconstruction with single patellar tunnel.
J Orthop Traumatol
January 2025
Department of Orthopedics, The First Affiliated Hospital of Wenzhou Medical University, Wenzhou, China.
Background: The objective of this review is to evaluate the methodological quality of meta-analyses and observe the consistency of the evidence they generated to provide comprehensive and reliable evidence for the clinical use of three-dimensional (3D) printing in surgical treatment of fracture.
Methods: We searched three databases (PubMed, Embase, and Web of Science) up until August 2024. The Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) standards were adhered to in this review.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!