Background: Tumor prostheses of the distal femur after revision surgery is associated with high rates of aseptic loosening, which has introduced great challenges to the survival of patients, but only a few studies have evaluated their X-ray imaging. The purpose of this study was to analyze the risk factors for recurrence of aseptic loosening and make recommendations to reduce the incidence of aseptic loosening after revision surgery of tumor prostheses in the distal femur.
Method: A retrospective analysis was performed on 23 patients who had revision surgery for distal femur prostheses due to aseptic loosening between June 2002 and June 2021. They were divided into two groups based on the condition of the prostheses after revision surgery: loosening group (9 patients) and control group (14 patients). Following the initial replacement, the length and diameter of the prosthetic intramedullary stem were measured through the standard full-length anteroposterior X-ray imaging of both lower limbs. The osteotomy length, femoral length and diameter, femoral intramedullary stem diameter, hip-knee-ankle angle (HKAA), mechanical lateral distal femoral angle (mLDFA), mechanical medial proximal tibial angle (mMPTA), and so on were measured as well. Following that, statistical analysis was performed.
Results: Patients in the loosening group had statistically significant differences in the ratio of prostheses length to femur length (71.89 ± 6.62) and the ratio of intramedullary stem diameter to femoral diameter (25.50 ± 6.90) (P < 0.05), when compared to the control group. The HKAA (175.58 ± 2.78), mLDFA (94.42 ± 2.57), and the deviation angle between the lower limb alignment and the tibial prostheses force line (2.23 ± 1.09) in the loosening group were significantly different from those in the control group (P < 0.05) on postoperative radiographs of the entire length of the lower limbs. The lowest score in intramedullary manubrium I indicated less osteolysis, while the highest score in intramedullary manubrium III indicated the most serious osteolysis, and the difference was statistically significant (P < 0.05).
Conclusions: Our study suggests that the use of longer and thicker intramedullary stems can effectively decrease the occurrence of aseptic loosening. Additionally, it is important to avoid using the original prostheses and reconstruct the standard line of lower limb force to further reduce the incidence of aseptic loosening. It is crucial to closely monitor the distal segment of the intramedullary stem for osteolysis after surgery.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10230718 | PMC |
http://dx.doi.org/10.1186/s12957-023-03047-0 | DOI Listing |
J Arthroplasty
January 2025
Georgetown University School of Medicine, MedStar Georgetown University Hospital, Washington, D.C, USA. Electronic address:
Introduction: As the number of revision total knee arthroplasties (rTKA) continues to rise, there is increasing interest in the use of contemporary rotating hinge prostheses. These devices often incorporate porous cones to fill bone defects and enhance long-term fixation. This study evaluated the clinical and functional outcomes and survivorship in rTKA patients utilizing a rotating hinge prosthesis with flexible titanium (FT) cones, porous tantalum (PT) cones, or no cones.
View Article and Find Full Text PDFJ Clin Med
January 2025
Shoulder and Elbow Surgery, Schulthess Clinic Zürich, 8008 Zürich, Switzerland.
: Classical reverse shoulder arthroplasty (RSA) with a high neck-shaft angle (NSA) of 155° has shown satisfactory outcomes. However, newer RSA designs aim to improve results by modifying the stem design. This study evaluates the 5-year outcomes of a stem design featuring a rectangular metadiaphyseal fixation and a 135° NSA.
View Article and Find Full Text PDFDespite an increasing number of studies examining the effect of Single-Photon Emission Computed Tomography/ Computed Tomography (SPECT/CT) on improvement of diagnosis of aseptic loosening, there is still a great deal of uncertainty regarding its applicability in diagnostic algorithm. Therefore, in this meta-analysis, we aimed to investigate the diagnostic performance of SPECT/CT for identification of aseptic loosening in patients with persistent pain following the total knee arthroplasty (TKA) and total hip arthroplasty (THA). Electronic databases including Medline, Scopus, Web of Science, Cochrane library, and Embase were systematically searched for identifying relevant published studies from their inception to April 2023.
View Article and Find Full Text PDFZhongguo Gu Shang
January 2025
Department of Orthopaedics, Xuzhou Central Hospital, Xuzhou 221009, Jiangsu, China.
Objective: To investigate the clinical utility of novel of new hematological markers in the preoperative diagnosis of periprosthetic joint infection (PJI).
Methods: A retrospective analysis was conducted on a total of 149 patients who underwent revision of total hip arthroplasty (THA) or total knee arthroplasty (TKA) at a single center between January 2016 and June 2022, including 63 males and 86 females, aged from 47 to 93 years old with an average of (69.5±11.
Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi
January 2025
Department of Artificial Joint Revision, Luoyang Orthopedic-Traumatological Hospital of Henan Province (Henan Provincial Orthopedic Hospital), Zhengzhou Henan, 450003, P. R. China.
Objective: To analyze the short-term effectiveness and safety of personalized three-dimensional (3D) printed customized prostheses in severe Paprosky type Ⅲ acetabular bone defects.
Methods: A retrospective analysis was conducted on 8 patients with severe Paprosky type Ⅲ acetabular bone defects and met the selection criteria between January 2023 and June 2024. There were 3 males and 5 females, with an average age of 64.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!