Background: Traumatic hemorrhage guidelines include point-of-care viscoelastic tests as a standard of care. Quantra (Hemosonics) is a device based on sonic estimation of elasticity via resonance (SEER) sonorheometry to assess whole blood clot formation.
Objectives: Our study aimed to assess the ability of an early SEER evaluation to detect blood coagulation test abnormalities in trauma patients.
Methods: We conducted an observational retrospective cohort study with data collected at hospital admission of consecutive multiple trauma patients from September 2020 to February 2022 at a regional level 1 trauma center. We performed a receiving operator characteristic curve analysis to determine the ability of the SEER device to detect blood coagulation test abnormalities. Four values on the SEER device were analyzed: clot formation time, clot stiffness (CS), platelet contribution to CS, and fibrinogen contribution to CS.
Results: A total of 156 trauma patients were analyzed. The clot formation time value predicted an activated partial thromboplastin time ratio of >1.5 with an area under the curve (AUC) of 0.93 (95% CI, 0.86-0.99). The AUC of the CS value in detecting an international normalized ratio of prothrombin time of >1.5 was 0.87 (95% CI, 0.79-0.95). The AUC of fibrinogen contribution to CS to detect a fibrinogen concentration of <1.5 g/L was 0.87 (95% CI, 0.80-0.94). The AUC of platelet contribution to CS to detect a platelet concentration of <50 G/L was 0.99 (95% CI, 0.99-1.00).
Conclusion: Our results suggest that the SEER device may be useful for the detection of blood coagulation test abnormalities at trauma admission.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10208882 | PMC |
http://dx.doi.org/10.1016/j.rpth.2023.100163 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!