Aim: Dose calculation accuracy between Anisotropic Analytical Algorithm (AAA) and Acuros XB (AXB) for various megavoltage (MV) photon beams for both flattening filter (FF) and flattening filter free (FFF) beams and to validate the accuracy of these dose calculations using inhomogeneous phantom in volumetric modulated arc therapy (VMAT).
Material And Methods: A Cheese Phantom having 20 holes that can be filled with all virtual water plugs or set of density calibration plugs was used for VMAT planning using two different algorithms using either single or double arc. Further phantom was used irradiate plan in linear accelerator and the point doses measured using a 0.053 cc A1SL ionization chamber along electrometer . Different plans, cylindrical shape, C-shaped and donut targets were planned 6MV, 10MV, 6FFF MV and 10FFF MV beam energy.
Result: The minimum average mean dose difference was 1.2% for PTV structures between AAA and AXB (p=0.02). Apart from these structures, the following density plugs have a more than 2% difference in maximum dose with statistical significance. (i) Solid water (MD=6.1%, p=0.016), (ii) Bone 200 (2.3%, p=0.029), (iii) CB_30% (MD=2.4%, p=0.050) and (iv) Cortical bone (MD=4.3%, p=0.018). In 6MV FFF and 10 MV FFF plans, the difference between AAA and AXB was not statistically significant (Fig 3). The Conformity index for the AAA less than that of AXB, in all energies and for all the PTVs. The CI was better in AXB than AAA, but the CI was not having much variation due to changes in beam energies, particularly for Cylinder shaped PTV.
Conclusion: All combinations of beam energy AAA showed higher values in the maximum dose than the Acuros XB, except for the lung insert. Nonetheless, AAA showed a higher mean dose than the Acuros XB. Differences between these two algorithms for most of the beam energies are minimal.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10495917 | PMC |
http://dx.doi.org/10.31557/APJCP.2023.24.5.1677 | DOI Listing |
Phys Eng Sci Med
December 2024
Icon Cancer Centre, Windsor Gardens, SA, Australia.
High-density materials used for dental restorations are poorly defined in CT imaging due to scanner limitations. Studies have established that Eclipse offers poor agreement with delivered dose in situations involving high-density material. Defining the accuracy of dose algorithms in situations involving high-density overrides would improve clinical outcomes both for target coverage and OAR sparing.
View Article and Find Full Text PDFJ Appl Clin Med Phys
August 2024
BC Cancer, Vancouver, British Columbia, Canada.
Background And Purpose: This study recommends clinical epidermal dose calculation methods based on in-vivo film measurements and registered skin dose distributions with the Eclipse (Varian Medical Systems) treatment planning system's Analytical Anisotropic Algorithm (AAA) and Acuros XB (AXB) dose calculation algorithms.
Materials And Methods: Eighteen AAA V13.6 breast plans were recalculated using AXB (dose to medium) V13.
Asian Pac J Cancer Prev
May 2024
Department of Physics, School of Advanced Sciences, Vellore Institute of Technology, Vellore, Tamil Nadu, India.
Aim: To assess the precision of dose calculations for Volumetric Modulated Arc Therapy (VMAT) using megavoltage (MV) photon beams, we validated the accuracy of two algorithms: AUROS XB and Analytical Anisotropic Algorithm (AAA). This validation will encompass both flattening filter (FF) and flattening filter-free beam (FFF) modes, using AAPM Medical Physics Practice Guideline (MPPG 5b).
Materials And Methods: VMAT validation tests were generated for 6 MV FF and 6 MV FFF beams using the AAA and AXB algorithms in the Eclipse V.
J Appl Clin Med Phys
July 2024
Department of Radiation Oncology and Image-Applied Therapy, Kyoto University Graduate School of Medicine, Kyoto, Japan.
Background: For patient-specific quality assurance (PSQA) for small targets, the dose resolution can change depending on the characteristics of the dose calculation algorithms.
Purpose: This study aimed to evaluate the influence of the dose calculation algorithms Acuros XB (AXB), anisotropic analytical algorithm (AAA), photon Monte Carlo (pMC), and collapsed cone (CC) on a helical diode array using volumetric-modulated arc therapy (VMAT) for small targets.
Materials And Methods: ArcCHECK detectors were inserted with a physical depth of 2.
Med Dosim
August 2024
Department of Radiotherapy, Baheya center for early detection and treatment of breast cancer, Giza, Egypt; Department of Radiotherapy, National cancer institute, Cairo University, Cairo, Egypt.
This study aimed to determine the dosimetric value of flattening filter-free (FFF) beams compared to flattening filter (FF) beams using different algorithms in the treatment planning of thoracic spine stereotactic body radiation therapy (SBRT). A total of 120 plans were created for 15 patients using the Anisotropic Analytical Algorithm (AAA) and the Acuros External Beam (AXB) algorithm with FF and FFF beams at 6 MV and 10 MV energies. Various dosimetric parameters were evaluated, including target coverage, dose spillage, and organs-at-risk sparing of the spinal cord and esophagus.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!