In this study, the impact of reconstruction sharpness on the visualization of the appendicular skeleton in ultrahigh-resolution (UHR) photon-counting detector (PCD) CT was investigated. Sixteen cadaveric extremities (eight fractured) were examined with a standardized 120 kVp scan protocol (CTDI 10 mGy). Images were reconstructed with the sharpest non-UHR kernel (Br76) and all available UHR kernels (Br80 to Br96). Seven radiologists evaluated image quality and fracture assessability. Interrater agreement was assessed with the intraclass correlation coefficient. For quantitative comparisons, signal-to-noise-ratios (SNRs) were calculated. Subjective image quality was best for Br84 (median 1, interquartile range 1-3; ≤ 0.003). Regarding fracture assessability, no significant difference was ascertained between Br76, Br80 and Br84 ( > 0.999), with inferior ratings for all sharper kernels ( < 0.001). Interrater agreement for image quality (0.795, 0.732-0.848; < 0.001) and fracture assessability (0.880; 0.842-0.911; < 0.001) was good. SNR was highest for Br76 (3.4, 3.0-3.9) with no significant difference to Br80 and Br84 ( > 0.999). Br76 and Br80 produced higher SNRs than all kernels sharper than Br84 ( ≤ 0.026). In conclusion, PCD-CT reconstructions with a moderate UHR kernel offer superior image quality for visualizing the appendicular skeleton. Fracture assessability benefits from sharp non-UHR and moderate UHR kernels, while ultra-sharp reconstructions incur augmented image noise.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10217602PMC
http://dx.doi.org/10.3390/diagnostics13101677DOI Listing

Publication Analysis

Top Keywords

image quality
16
fracture assessability
16
appendicular skeleton
8
uhr kernels
8
interrater agreement
8
br76 br80
8
br80 br84
8
br84 0999
8
moderate uhr
8
fracture
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!