Background: Whether deep learning-based CT reconstruction could improve lesion conspicuity on abdominal CT when the radiation dose is reduced is controversial.

Objectives: To determine whether DLIR can provide better image quality and reduce radiation dose in contrast-enhanced abdominal CT compared with the second generation of adaptive statistical iterative reconstruction (ASiR-V).

Aims: This study aims to determine whether deep-learning image reconstruction (DLIR) can improve image quality.

Method: In this retrospective study, a total of 102 patients were included, who underwent abdominal CT using a DLIR-equipped 256-row scanner and routine CT of the same protocol on the same vendor's 64-row scanner within four months. The CT data from the 256-row scanner were reconstructed into ASiR-V with three blending levels (AV30, AV60, and AV100), and DLIR images with three strength levels (DLIR-L, DLIR-M, and DLIR-H). The routine CT data were reconstructed into AV30, AV60, and AV100. The contrast-to-noise ratio (CNR) of the liver, overall image quality, subjective noise, lesion conspicuity, and plasticity in the portal venous phase (PVP) of ASiR-V from both scanners and DLIR were compared.

Results: The mean effective radiation dose of PVP of the 256-row scanner was significantly lower than that of the routine CT (6.3±2.0 mSv vs. 2.4±0.6 mSv; p< 0.001). The mean CNR, image quality, subjective noise, and lesion conspicuity of ASiR-V images of the 256-row scanner were significantly lower than those of ASiR-V images at the same blending factor of routine CT, but significantly improved with DLIR algorithms. DLIR-H showed higher CNR, better image quality, and subjective noise than AV30 from routine CT, whereas plasticity was significantly better for AV30.

Conclusion: DLIR can be used for improving image quality and reducing radiation dose in abdominal CT, compared with ASIR-V.

Download full-text PDF

Source
http://dx.doi.org/10.2174/1573405620666230525104809DOI Listing

Publication Analysis

Top Keywords

image quality
24
radiation dose
16
256-row scanner
16
lesion conspicuity
12
quality subjective
12
subjective noise
12
image
9
image reconstruction
8
compared second
8
second generation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!