Optimization of chest CT protocols based on pixel image matrix, kernels and iterative reconstruction levels - A phantom study.

Radiography (Lond)

Faculty of Health Sciences, Department of Life Sciences and Health, Oslo Metropolitan University, Pilestredet 48, 0130 Oslo, Norway. Electronic address:

Published: July 2023

Introduction: This study investigated the impact of high matrix image reconstruction in combination with different reconstruction kernels and levels of iterative reconstructions on image quality in chest CT.

Methods: An anthropomorphic chest phantom (Kyoto Kagaku Co., Ltd., Kyoto, Japan), and a Catphan® 600 (The Phantom Laboratory, Greenwich, NY, USA) phantom were scanned using a dual source scanner. Standard institutional protocol with 512 × 512 matrix was used as a reference. Reconstructions were performed for 768 × 768 and 1024 × 1024 matrices and all possible combinations of three different kernels and five levels of iterative reconstructions were included. Signal difference to noise ratio (SdNR) and line pairs per cm (lp/cm) were manually measured. A Linear regression model was applied for objective image analysis (SdNR) and inter-and intra-reader agreement was given as Cohen's kappa for the visual image assessment.

Results: Matrix size did not have a significant impact on SdNR (p = 0.595). Kernel (p = 0.014) and ADMIRE level (p = 0.001) had a statistically significant impact on SdNR. The spatial resolution ranged from 7 lp/cm to 9 lp/cm. The highest spatial resolution was achieved using kernel Br64 and ADMIRE 1, 2 and 3 in both 768- and 1024-matrices, and with Br59 with ADMIRE 2 and 4 and 768-matrix, all visualizing 9 lp/cm. Both readers scored kernel Br59 highest, and the scoring increased with increasing levels of Iterative Reconstruction.

Conclusion: Matrix size did not influence image quality, however, the choice of kernel and degree of IR had an impact on objective and visual image quality in 768 - and 1024-matrices, suggesting that increased degree of IR may improve diagnostic image quality in chest CT.

Implications For Practice: Image quality in CT of the lung may be improved by increasing the level of IR.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.radi.2023.05.005DOI Listing

Publication Analysis

Top Keywords

image quality
20
levels iterative
12
image
9
kernels levels
8
iterative reconstructions
8
quality chest
8
visual image
8
matrix size
8
impact sdnr
8
spatial resolution
8

Similar Publications

Multi-modal systems extract information about the environment using specialized sensors that are optimized based on the wavelength of the phenomenology and material interactions. To maximize the entropy, complementary systems operating in regions of non-overlapping wavelengths are optimal. VIS-IR (Visible-Infrared) systems have been at the forefront of multi-modal fusion research and are used extensively to represent information in all-day all-weather applications.

View Article and Find Full Text PDF

Optimizing Correction Factors on Color Differences for Automotive Painting Services.

Sensors (Basel)

December 2024

Department of Management and Industrial Engineering, University of Petrosani, 332003 Petrosani, Romania.

Currently, the automotive sector is showing increased demands regarding the color of cars in general, but especially the quality and the time of painting, in particular. Companies working in this industry, especially in specialized painting services, must perform work of impeccable quality in the shortest possible time in order to be efficient. Color differences that appear in different areas of the car result from the use of different formulas for obtaining color.

View Article and Find Full Text PDF

A WSN and LoRa Hybrid Multimedia Transmission Protocol for Scalar Data and Image Transmission.

Sensors (Basel)

December 2024

Ubicom Laboratory, Department of Electrical, Electronic and Computer Engineering, University of Ulsan, Ulsan 44610, Republic of Korea.

The proposed protocol features reliable and fast image transmission while periodically transmitting scalar data without interruption by allowing two networks, a LoRa network and a wireless sensor network, with different transmission characteristics to cooperate. It adopts the RT-LoRa protocol for periodic scalar data transmission and uses a WSN-based pipelined transmission method that leverages single-hop message transmission of a LoRa network for image transmission. Thus, it can not only eliminate the control message overhead for time synchronization, slot scheduling, and path establishment for pipelined image transmission in WSNs but also eliminate interferences within WSNs, such as data collisions and data and message collisions, during pipelined image transmission, thereby enabling high reliability and fast transmission.

View Article and Find Full Text PDF

To achieve high-precision 3D reconstruction, a comprehensive improvement has been made to the binocular structured light calibration method. During the calibration process, the calibration object's imaging quality and the camera parameters' nonlinear optimization effect directly affect the caibration accuracy. Firstly, to address the issue of poor imaging quality of the calibration object under tilted conditions, a pixel-level adaptive fill light method was designed using the programmable light intensity feature of the structured light projector, allowing the calibration object to receive uniform lighting and thus improve the quality of the captured images.

View Article and Find Full Text PDF

Advancements in Raman light sheet microscopy have provided a powerful, non-invasive, marker-free method for imaging complex 3D biological structures, such as cell cultures and spheroids. By combining 3D tomograms made by Rayleigh scattering, Raman scattering, and fluorescence detection, this modality captures complementary spatial and molecular data, critical for biomedical research, histology, and drug discovery. Despite its capabilities, Raman light sheet microscopy faces inherent limitations, including low signal intensity, high noise levels, and restricted spatial resolution, which impede the visualization of fine subcellular structures.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!