Here, we have developed and applied a triple-barrel microelectrode. This device incorporates a platinum disk working electrode, a platinum disk counter electrode, and a low-leakage Ag/AgCl reference electrode into a small probe. We demonstrate that the incorporated low-leakage reference electrode shows similar voltammetry, potentiometry, and drift when compared to a commercial reference electrode in bulk solution. We also demonstrate the versatility of such a small three-channel system via voltammetry in nanoliter droplets and through electroanalysis of captured aerosols. Finally, we demonstrate the probe's potential utility in single-cell electroanalysis by making measurements within salmon eggs.