A Dual-Chamber Leadless Pacemaker.

N Engl J Med

From Amsterdam University Medical Centers, Amsterdam (R.E.K., K.T.N.B.); Icahn School of Medicine at Mount Sinai (V.Y.R.) and Weill Cornell Medicine-New York Presbyterian Hospital (J.E.I.) - both in New York; HonorHealth Cardiac Arrhythmia Group, Scottsdale, AZ (R.D.); Foothills Medical Centre, Calgary, AB, Canada (D.V.E.); Centre Hospitalier Régional Universitaire Albert Michallon, Grenoble, France (P.D.); Texas Cardiac Arrhythmia Institute, Austin (R.C.); Azienda Ospedaliero-Universitaria Pisana, Pisa, Italy (M.G.B.); Tokyo Women's Medical University, Tokyo (M.S.); Deutsches Herzzentrum der Charité, Berlin (G.H.); Na Homolce Hospital, Prague, Czech Republic (V.Y.R., P.N.); Huntington Memorial Hospital, Pasadena (M.R.), and Abbott Medical, Sylmar (J.R.N., L.G., C.H.) - both in California; and the Cleveland Clinic, Cleveland (D.J.C.).

Published: June 2023

AI Article Synopsis

  • A dual-chamber leadless pacemaker system, involving devices placed in both the right atrium and right ventricle, aims to improve treatment options for patients needing pacemakers without the conventional wiring.
  • In a study of 300 patients, the new system demonstrated high safety, with 90.3% of participants avoiding serious complications within 90 days, which surpassed the target benchmark.
  • The effectiveness of the pacemaker was also confirmed, as 90.2% of patients achieved satisfactory atrial capture and sensing performance, indicating it can reliably support necessary heart functions.

Article Abstract

Background: Single-chamber ventricular leadless pacemakers do not support atrial pacing or consistent atrioventricular synchrony. A dual-chamber leadless pacemaker system consisting of two devices implanted percutaneously, one in the right atrium and one in the right ventricle, would make leadless pacemaker therapy a treatment option for a wider range of indications.

Methods: We conducted a prospective, multicenter, single-group study to evaluate the safety and performance of a dual-chamber leadless pacemaker system. Patients with a conventional indication for dual-chamber pacing were eligible for participation. The primary safety end point was freedom from complications (i.e., device- or procedure-related serious adverse events) at 90 days. The first primary performance end point was a combination of adequate atrial capture threshold and sensing amplitude at 3 months. The second primary performance end point was at least 70% atrioventricular synchrony at 3 months while the patient was sitting.

Results: Among the 300 patients enrolled, 190 (63.3%) had sinus-node dysfunction and 100 (33.3%) had atrioventricular block as the primary pacing indication. The implantation procedure was successful (i.e., two functioning leadless pacemakers were implanted and had established implant-to-implant communication) in 295 patients (98.3%). A total of 35 device- or procedure-related serious adverse events occurred in 29 patients. The primary safety end point was met in 271 patients (90.3%; 95% confidence interval [CI], 87.0 to 93.7), which exceeded the performance goal of 78% (P<0.001). The first primary performance end point was met in 90.2% of the patients (95% CI, 86.8 to 93.6), which exceeded the performance goal of 82.5% (P<0.001). The mean (±SD) atrial capture threshold was 0.82±0.70 V, and the mean P-wave amplitude was 3.58±1.88 mV. Of the 21 patients (7%) with a P-wave amplitude of less than 1.0 mV, none required device revision for inadequate sensing. At least 70% atrioventricular synchrony was achieved in 97.3% of the patients (95% CI, 95.4 to 99.3), which exceeded the performance goal of 83% (P<0.001).

Conclusions: The dual-chamber leadless pacemaker system met the primary safety end point and provided atrial pacing and reliable atrioventricular synchrony for 3 months after implantation. (Funded by Abbott Medical; Aveir DR i2i ClinicalTrials.gov number, NCT05252702.).

Download full-text PDF

Source
http://dx.doi.org/10.1056/NEJMoa2300080DOI Listing

Publication Analysis

Top Keywords

leadless pacemaker
16
dual-chamber leadless
12
leadless pacemakers
8
atrioventricular synchrony
8
pacemaker system
8
primary safety
8
safety point
8
device- procedure-related
8
procedure-related serious
8
serious adverse
8

Similar Publications

A rare cardiac presentation of a lymphoma: case report.

Eur Heart J Case Rep

January 2025

Department of Cardiology, Klinik Floridsdorf, Brünnerstraße 68, Vienna 1210, Austria.

Background: Cardiac lymphoma is a rare disease that can present in various ways. Additionally, atypical clinical presentation makes the diagnosis even more challenging. The most common type of cardiac lymphoma is diffuse large B-cell lymphoma.

View Article and Find Full Text PDF

Minimizing for Maximum Benefit: An Illustrative Case-Series of Atrial Only Leadless Pacing.

Pacing Clin Electrophysiol

December 2024

Division of Cardiology, Department of Medicine, Weill Cornell Medicine, New York Presbyterian Hospital, New York, USA.

Leadless pacing technology now includesdedicated atrial helix-fixation leadless pacemakers (LPs), expanding theapplication of leadless devices for patients with sinus node dysfunction andatrioventricular block during sinus rhythm. This first reportedcase-series of atrial LPs describes and discusses the potential use-casescenarios of recently approved helix-fixation atrial LPs. The article highlights important concepts regarding their use, including implantationtechniques, programming, battery conservation, and the low rate of progressionof AV block in patients implanted with AAI(R) pacemakers.

View Article and Find Full Text PDF

Reel's syndrome (RS) is an unusual cause of pacemaker lead dislodgement. We present the case of a 59-year-old female patient with Down syndrome (DS) implanted with a dual-chamber endovascular pacemaker due to symptomatic sinus node disfunction, reporting several syncopal episodes in last days and showing abnormal electrical parameters at the 2-months follow-up due to RS. The malfunctioning device was removed and an endocardial leadless pacing system was implanted.

View Article and Find Full Text PDF

Background: Delayed lead perforation is a rare complication of cardiac implantable electronic device (CIED). Clinical presentations range from completely asymptomatic to pericardial tamponade. Surgical lead extraction is recommended and transvenous lead extraction (TLE) with surgical backup is an alternative method.

View Article and Find Full Text PDF

Cardiac Implantable Electronic Devices in Cardiac Transplant Patients: A Comprehensive Review.

Cardiol Rev

November 2024

Department of Cardiovascular Medicine, Cardiovascular Analytics Group, Islamabad, Pakistan.

A fraction of patients (approximately 10%) undergoing heart transplantation require permanent pacemaker (PPM) implantation due to sinus node dysfunction or atrioventricular block, occurring either shortly after surgery or later. The incidence of PPM implantation has declined to less than 5% with the introduction of bicaval anastomosis transplantation surgery. Pacing dependency during follow-up varies among recipients.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!