Background: Dual-mobility constructs have gained popularity to mitigate dislocations after high-risk primary and revision total hip arthroplasties. Contemporary data have indicated that malseating of modular dual-mobility liners occurs in up to 6% of cases. The purpose of this cadaveric-based radiographic study was to determine the ability to accurately determine if modular dual-mobility liners were seated.
Methods: There were 10 hips (5 cadaveric pelvic specimens) used to implant modular dual-mobility liners of 2 designs. One had a liner that seated flush and the other had an extended rim. There were 20 constructs that were well-seated and 20 constructs were intentionally malseated. A comprehensive series of radiographs was reviewed by 2 blinded surgeons. Statistical analyses included Chi-squared testing, logistic regressions, and kappa statistics.
Results: Radiographic assessment of liner malseating was not accurate with an elevated rim design with misdiagnosis in 40% (16 of 40). The flush design had diagnostic errors in 5% (2 of 40; P = .0002). Logistic regressions demonstrated a significantly higher risk of misdiagnosing a malseated liner in the elevated rim group (odds ratio 13). There were 12 of 16 misdiagnoses in the elevated rim group failing to recognize a malseated liner. Each surgeon had almost perfect agreement for intraobserver reliability for flush designs (k 0.90) and fair agreement in the elevated rim design (k 0.35).
Conclusion: A comprehensive series of plain radiographs can reliably detect a malseated modular dual-mobility liner with a flush rim design in 95% of cases. However, elevated rim designs are more difficult to accurately identify malseating on plain radiographs.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10704932 | PMC |
http://dx.doi.org/10.1016/j.arth.2023.05.020 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!